![Unipulse F701+ Скачать руководство пользователя страница 140](http://html.mh-extra.com/html/unipulse/f701/f701_operation-manual_843485140.webp)
133
8
List of Setting Items
133
Li
st
o
f S
et
tin
g
Ite
m
s
Chapter
8
7 Extended function selection 1
00000 00000 to 91111
◎
W48(R/W)
BCD data update rate
0
(A/D conversion rate
100 times/sec)
6: Once/sec
5: Twice/sec
4: 5 times/sec
3: 10 times/sec
2: 20 times/sec
1: 50 times/sec
0: 100 times/sec
(A/D conversion rate
500 times/sec)
7: Once/sec
6: Twice/sec
5: 5 times/sec
4: 10 times/sec
3: 20 times/sec
2: 50 times/sec
1: 100 times/sec
0: 500 times/sec
Filter in stable condition
0
(A/D conversion rate
100 times/sec)
1: Do not insert
0: Insert (64 times)
(A/D conversion rate
500 times/sec)
1: Do not insert
0: Insert (256 times)
Motion detect
0
1: Checker mode
0: Stable mode
Gross weight/net weight
display switch
0
1: External input mode
0: Internal, external mode
During discharge control
0
1: Net weight displayed
without sign reverse
0: Net weight displayed with
sign reverse
8
Moving average filter
030
(When A/D conversion rate
100 times/sec):
1 to 128 times
(When A/D conversion rate
500 times/sec):
1 to 512 times
◎
W39(R/W)
9 Measurement Act compliance
0000 0000 to 1111
◎
W49(R/W)
One-touch tare subtraction
0
1: Accept only when stable
0: Accept regularly
Tare subtraction range
0
1: 0 < tare
≦
capacity
0: Total range
Prohibits ON/OFF of tare
setting and preset tare
weight when one-touch tare
subtraction is enabled
0
1: Enabled
0: Disabled
Tare weight displayed with
[0] key
0
1: Enabled 0: Disabled