6.FUNCTIONS
63
6-9. Analog Filter
This is a low-pass filter to eliminate noise components from input signals from the load cell, and
also stabilize the weight value by averaging.
Select the cutoff frequency from 2Hz, 4Hz, 6Hz and 8Hz.
By lowering the cutoff frequency, the display becomes more stable, but the response worsens.
On the contrary, the higher the frequency, the better the response becomes, but the more difficult
it is to make the display stable.
Setting / 3 : 8Hz 2 : 6Hz 1 : 4Hz 0 : 2Hz
CC-Link
0
35
MSB
LSB
RWw0006
RWw0005
RWw0004
Write the code No. and set value.
H
L
0
0 to 3
Code No.
Set value
remote register
Word size
Console
1) Select code 35.
2) Press the ENT key, and then set the
analog filter.
3) Determine the setting with the ENT key.
Analog filter