![Unipulse F156 Скачать руководство пользователя страница 45](http://html2.mh-extra.com/html/unipulse/f156/f156_operation-manual_3889947045.webp)
4.CALIBRATION
36
4. CALIBRATION
4-1. What is Calibration?
Operation of matching the F156 and load cell is called “calibration.” To be specific, adjustments
are made so that the F156 can exactly display 100.00kg when an actual load (or weight) of
100kg is put on the load cell (scale section) of the weighing machine to which the F156 is
connected for example. This operation is called actual load calibration.
Содержание F156
Страница 1: ...F156 WEIGHING CONTROLLER 07 JUN 2012 REV 1 40 OPERATION MANUAL CC Link Ver 1 10...
Страница 115: ...7 REMOTE CONSOLE UNIT 106 Monitor change Change the display with the key...
Страница 120: ...7 REMOTE CONSOLE UNIT 111 Monitor change Change the display with the keys 0 to 9 1 2 3 4 5 0...
Страница 155: ...13 DIMENSIONS 146 13 DIMENSIONS...
Страница 156: ...14 BLOCK DIAGRAM 147 14 BLOCK DIAGRAM...
Страница 160: ...Unipulse Corporation 9 11 Nihonbashi Hisamatsucho Chuo ku Tokyo 103 0005 Tel 81 3 3639 6120 Fax 81 3 3639 6130...