Information related to designated use
User Manual CWD2005 Plus
3. Information related to designated use
The analyzer should be placed as close as possible to the sampling location to avoid
unnecessary time delays.
Install the analyzer in a climatically stable area; see the manual of the analyzer.
The analyzer is not suitable for outdoor installation.
The exhaust gas must be led to the outside of the room
Danger!
The exhaust gas shall be dispose at safe location outside.
It may be possible that the exhaust gas contains toxic parts.
Follow exactly to the procedure described for handling the analyzer.
3.1 Foreseeable misuse
Each unintended use (or misuse) of the analyzer can impair the function of the analyzer, thus
establishing a direct threat to humans and the whole system.
Misuse will result in immediate loss of any liability claim.
Misuse occurs if:
-
neglecting the warnings attached to the analyzer
-
neglecting the procedure of commissioning and operation
-
operating the analyzer without purge air
-
operating the analyzer without a safety switch
-
disregarding the error signal
-
disregarding of the conditions of issue
-
neglecting the classified area
-
neglecting the voltage regulation of power supply on-site
The list above is not final. Please note the similar situations and conditions that deviate from
the intended application.
Warning!
Ignoring the safety references and the instruction can possibly cause an
escaping gas to a combustible mixture inside of the instrument.
It may cause explosion!
9