8.5 Datalogger commands
Table 8.34 –
continued from previous page
Column
Example value
Description
22-23
Carriage return (0x0d), Line feed (0x0a).
An example response is:
La,000,128,128,128,001
0123456789 123456789 123
8.5.17 Mutual access
Allows the SQM-LU-DL to log to its FLASH memory database while connected to a PC as well as while connected to
the battery pack.
The default is to only allow triggered logging while connected to the battery pack where there are no data wires (only
power wires) connected to the USB jack.
This mutual access settings allows logging while also connected to a PC (when data wires are connected to the USB
jack).
The Mutual access mode is set with the “
LDnx
” command, where n is 0 or 1. The Mutual access mode is read with the
“
Ldx
” command. The response of both setting and getting the mutual access value is shown in table 8.35:
Table 8.35: Response of setting or getting the mutual access mode
Column
Example value
Description
0-2
Ld,
Status mutual access.
3
1
0 = battery only logging is allowed.
1= PC and battery logging is allowed.
4-5
Carriage return (0x0d), Line feed (0x0a).
An example response is:
LD,1
012345
Unihedron SQM-LU-DL Operator’s Manual -
20220119