13 Troubleshooting
Table 13.10 –
continued from previous page
Solution
The following steps can be used to clean up the COM port entries in Windows:
1. Unplug all FTDI devices.
2. In the Windows device manager, tag ”show hidden devices”.
3. Delete the COM ports, but keep the driver.
4. Re-connect one FTDI device after the other.
Table 13.11: Meter cannot be found by Windows operating system
Problem
The windows computer cannot identify the SQM-LU-DL, and re-installing the FTDI device drivers
does not help.
Cause
The Windows FTDI driver installation may be storing old attempts at installing drivers that should
have been cleared out.
Solution
The following steps can be used to clean up the FTDI entries in Windows:
1. Install this FTDI “CDM Uninstaller” utility at:
www.ftdichip.com/Support/Utilities/CDMUninstaller v1.4.zip
2. The detailed instructions for its operation are here:
www.ftdichip.com/Support/Utilities/CDM Uninst GUI Readme.html
3. Here is a quick summary of the instructions:
4. Start up the CDM uninstaller program
5. Press - Add
6. Press - Remove devices
7. Wait 5 minutes or so until complete
8. Close
9. Re-insert SQM-LU device
10. Wait 5 minutes or so for Windows to install driver and port
Table 13.12: Cannot get a reading
Problem
Cannot get a reading
Cause
Driver is not installed or the SQM-LU-DL is not connected
Table continued on next page ...
98
Unihedron SQM-LU-DL Operator’s Manual -
20220119
Содержание SQM-LU-DL
Страница 2: ......
Страница 17: ...17...
Страница 19: ...Figure 4 1 Front and back of of unit Unihedron SQM LU DL Operator s Manual 20220119 19...