Precision Water Level Instrument
Unidata Manual 6293 - Precision Water Level Instruments Issue 2.1
15
4.3 Connecting to an External Data Logger via HSIO
Removing the front cover reveals a termination block to which the data logger cable
is connected. This cable exits through the gland in the side of the instrument.
External power may be connected to the designated terminals.
Precision Water Level Instrument
STARLOGGER Prologger
Wire
Colour
Logger Function
Pin Term
Pin Term
Red
+5.00V DC sync (terminal 3)
13
15, 16 or 17 18 14
Green (terminal
7)
23 14
28
17
Blue
DATA IN (s0) (terminal 5)
21
11
32 11
White
DATA OUT (terminal 6)
Yellow Serial CLOCK (terminal 4)
9
12
14 12
Notes
Pin # refers to the pins of the INPUT SIGNALS connector on a logger.
Term # refers to numbered screw terminals on a Field Termination Strip. This is the
connection method you are most likely to use.
1
Refer to the supplement included with your logger or Field Termination Strip for
more information about these connections.