background image

 
 


 

B.

 

Sample Preparation 

1.        Make a blank reference solution by filling a clean square cuvette (or test tube)                
half full with distilled/de-ionized water or other specified solvent.  Wipe the cuvette with tissue 
to remove the fingerprints or droplets of liquid. 
2.        Insert the blank square cuvette into the square cuvette adapter and place the                  
adapter in the Sample Compartment (if using test tube, insert it directly into the Sample 
Compartment, aligning the guide mark (if present) with the guide mark at the front of the 
Sample Compartment).  Close the lid. 
3.        Set 0.000A or 100%T with the 

0A/100%T 

button.   

 

NOTE

:  This step fulfills the instructions on the front of the spectrophotometer  

 
4.

 

Remove the blank square cuvette or test tube.  Set it aside in case you may need to 
adjust the 0A/100%T using 

0A/100%T 

button later (i.e. change the wavelength). 

 

C.

 

Sample Analysis 

1. 

Rinse a second square cuvette with a small amount of the sample solution to be tested.  
Fill the square cuvette half full and wipe clean the exterior of the cuvette. 

2.        Put the sample cuvette in the Sample Compartment.  Close the lid. 
3.        Read the 

T

 or 

A

 from the Digital Display window.  Remove the sample cuvette or            

test tube. 
4.        If you are to test the same sample at other wavelengths, repeat 

step 3 to B. 1 

             4 

for each wavelength. 
5.        For each new sample you analyze, repeat 

step 2 to C. 4

 

Output and Data Processing 

 

USB Data Port 

Model 

1100

 has a USB port which will allow you to interface with a Windows

®

 based computer 

for use with optional software, contact UNICO for details. 
The

 Spectrophotometer Application Software 

is Windows-based software designed to operate 

with your spectrophotometer using the USB port.  

 

The software runs on a PC with Windows

®

 

XP/7/10 operating system installed.  You can export data files collected from the 
spectrophotometer into Excel

®

 format if needed.  The software offers two additional analytical 

methods: 

Standard Curve 

and 

Absorbance vs. Time Kinetics

.  It

 

performs the following 

methods for analysis: 

 

Absorbance/%Transmittance/Concentration

 Measure the Absorbance, 

%Transmittance, Concentration/Standard, or Concentration/Factor at a single wavelength 
within the range of 

335

-

1000

 

nm.

 

 

Standard Curve

 Create a calibration curve (choice of 4 curve fits) with up to 8 standard 

solutions at a single wavelength to determine concentrations of unknown samples. 

 

Absorbance vs. Time Kinetics

 M

easure a sample’s absorbance change over a selected 

period of time, store the results in a data table, and display the results graphically. 

Содержание S-1100

Страница 1: ...SPECTROPHOTOMETER MODEL S 1100 Operation Manual United Products Instruments Inc Dayton NJ USA www unicosci com 732 274 1155...

Страница 2: ...trophotometer make sure it is set to the voltage of the local power supply see Installation section The main plug shall be inserted in a socket provided with a protective earth grounded contact The pr...

Страница 3: ...e Monochromator Sample Compartment Detector Digital Display The light from the Halogen Lamp is focused on the entrance slit of the Monochromator where the collimating mirror directs the beam onto the...

Страница 4: ...ion Place the instrument on a dry dust free lab bench top In order to have the best performance from your instrument keep it as far as possible from any strong magnetic or electrical fields or any ele...

Страница 5: ...ept round tube or square cuvets Follow the steps below to change the Cuvette Sample Holder Open the lid of the Sample Holder and locate the Sample Holder Locking Screw Use the Allen Wrench included to...

Страница 6: ...sample cuvette or test tube 4 If you are to test the same sample at other wavelengths repeat step 3 to B 1 4 for each wavelength 5 For each new sample you analyze repeat step 2 to C 4 Output and Data...

Страница 7: ...llowing methods to check wavelength calibration Please note that this test requires the use of a didymium filter or a holmium oxide filter Contact Carolina for availability and pricing of these filter...

Страница 8: ...uld be within 805 809 nm The wavelength accuracy of the unit is 2 nm 9 If a middle wavelength check is desired set the wavelength to 522 nm optional 10 Make sure the cuvette adapter is empty and place...

Страница 9: ...rect reading for the stray light value Troubleshooting PROBLEM Possible Cause Solution Instrument Inoperative Power indicator has no light Power cord not connected to outlet Plug instrument in Dead Po...

Страница 10: ...d properly Check if lamp has been properly installed or has moved during transit Refer to lamp Replacement instructions in this manual Lamp old or defective Replace with a new lamp Sample Holder Misal...

Отзывы: