Introduction 

The UNICO 1100 Spectrophotometers are single beam, general purpose instruments designed 
to meet the needs of the conventional laboratory 
Featuring a digital display of photometric result, easy operation, and wavelength range of  
335 to 1000 nm, UNICO 1100 series are ideal for measurements in the visible wavelength 
region of the electromagnetic spectrum. 
 

Working Principle 

The spectrophotometer consists of five parts: 
1) 

Light Source

 (Halogen Lamp) to supply the light;  

2) 

A Monochromator

 to isolate the wavelength of interest and eliminate the unwanted second 

order radiation;  
3) 

A Sample Compartment

 to accommodate the sample solution;  

4) 

A Detector

 to receive and convert the transmitted light to an electrical signal;  

5) 

A Digital Display

 to show absorbance or transmittance. Diagram below illustrates the 

relationship between these parts.  
 
 
 
 
 Light Source       Monochromator     Sample Compartment         Detector             Digital Display 

 

The light from the Halogen Lamp is focused on the entrance slit of the Monochromator where 
the collimating mirror directs the beam onto the grating. The grating disperses the light beam 
to produce the spectrum, a portion of which is focused on the exit slit of the Monochromator 
by a collimating mirror. From here the beam is passed to a Sample Compartment through one 
of the filters, which helps to eliminate unwanted second order radiation from the diffraction 
grating. Upon leaving the Sample Compartment, the beam is passed to the silicon photodiode 
Detector and causes the Detector to produce an electrical signal that is shown on the Display 
 

Specifications

  

 

 
 
 
 
 
 
 
 
 
 
 
 

 

Model S-1100 

Wavelength Range 

335-1000 nm 

Spectral Bandpass 

20 nm 

Wavelength Accuracy 

± 2 nm

 

Wavelength Repeatability 

± 1 nm

 

Stray Radiant Energy 

< 0.5 %T at 340 and 400 nm 

Photometric Range 

0-125%T 0-2.0 Abs 

Photometric Accuracy 

± 2.0%T

 

Power Requirements 

115/230 V ±10%, 60/50 Hz

 

Dimensions 

408W X 308D X 185H (mm) 

Net Weight 

6kg (13 lbs) 

100%T 
  0 A 

Содержание S-1100

Страница 1: ...SPECTROPHOTOMETER MODEL S 1100 Operation Manual United Products Instruments Inc Dayton NJ USA www unicosci com 732 274 1155...

Страница 2: ...trophotometer make sure it is set to the voltage of the local power supply see Installation section The main plug shall be inserted in a socket provided with a protective earth grounded contact The pr...

Страница 3: ...e Monochromator Sample Compartment Detector Digital Display The light from the Halogen Lamp is focused on the entrance slit of the Monochromator where the collimating mirror directs the beam onto the...

Страница 4: ...ion Place the instrument on a dry dust free lab bench top In order to have the best performance from your instrument keep it as far as possible from any strong magnetic or electrical fields or any ele...

Страница 5: ...ept round tube or square cuvets Follow the steps below to change the Cuvette Sample Holder Open the lid of the Sample Holder and locate the Sample Holder Locking Screw Use the Allen Wrench included to...

Страница 6: ...sample cuvette or test tube 4 If you are to test the same sample at other wavelengths repeat step 3 to B 1 4 for each wavelength 5 For each new sample you analyze repeat step 2 to C 4 Output and Data...

Страница 7: ...llowing methods to check wavelength calibration Please note that this test requires the use of a didymium filter or a holmium oxide filter Contact Carolina for availability and pricing of these filter...

Страница 8: ...uld be within 805 809 nm The wavelength accuracy of the unit is 2 nm 9 If a middle wavelength check is desired set the wavelength to 522 nm optional 10 Make sure the cuvette adapter is empty and place...

Страница 9: ...rect reading for the stray light value Troubleshooting PROBLEM Possible Cause Solution Instrument Inoperative Power indicator has no light Power cord not connected to outlet Plug instrument in Dead Po...

Страница 10: ...d properly Check if lamp has been properly installed or has moved during transit Refer to lamp Replacement instructions in this manual Lamp old or defective Replace with a new lamp Sample Holder Misal...