1100 Spectrophotometer Operation Panel

 

 

Mode Indicator

Show the current measurement mode (

T

--%Transmittance, 

A

--Absorbance) 

 

MODE

 Button: 

Switch between 

and 

A

 measurement modes. 

 

0A/100%T

 Button: 

Adjust Digital Display reading to 100%T or 0.000A when blank reference 

solution is in Sample Compartment. 

Sample Compartment

Accepts 10 mm test tube or 10 mm square cuvette (the square cuvette 

adapter is required). 

WAVELENGTH

 (Wavelength Control) Knob: 

Select desired wavelength in nanometers (nm). 

Wavelength Readout

 Window:

 Display desired wavelength. 

 

Changing Sample Holders 

Your spectrophotometer comes standard with a 10 mm path length 

Sample Holder 

which will 

accept round tube, or square cuvets.   
Follow the steps below to change the Cuvette 

Sample Holder

 

Open the lid of the 

Sample Holder

 and locate the 

Sample Holder Locking Screw 

 

 

Use the 

Allen Wrench

 (included) to loosen the 

Screw

 counterclockwise 

 

Remove the 

Sample Holder 

you

 

want to change, insert the one you want to install, align 

it properly, and fasten the 

Screw

 

Screw

Allen Wrench

 

 
 

Basic Operation 

 
NOTE: 

OPERATING INSTRUCTIONS

 are printed on the front panel of the instrument. 

 
Sample Preparation and Analysis 
A. 

Spectrophotometer Warm-up

 

1. 

Turn on the spectrophotometer by turning on the Power Switch (

IO

).  Allow a minimum 

15 minutes for the instrument to warm up. 

2. 

Select either the 

T

 (Transmittance Mode) or 

A

 (Absorbance Mode) by pressing the 

MODE

 button until the red light for 

or 

A

 is on. 

3.

 

Select the desired wavelength by turning the 

WAVELENGTH

 control knob.   

 
 

Содержание S-1100

Страница 1: ...SPECTROPHOTOMETER MODEL S 1100 Operation Manual United Products Instruments Inc Dayton NJ USA www unicosci com 732 274 1155...

Страница 2: ...trophotometer make sure it is set to the voltage of the local power supply see Installation section The main plug shall be inserted in a socket provided with a protective earth grounded contact The pr...

Страница 3: ...e Monochromator Sample Compartment Detector Digital Display The light from the Halogen Lamp is focused on the entrance slit of the Monochromator where the collimating mirror directs the beam onto the...

Страница 4: ...ion Place the instrument on a dry dust free lab bench top In order to have the best performance from your instrument keep it as far as possible from any strong magnetic or electrical fields or any ele...

Страница 5: ...ept round tube or square cuvets Follow the steps below to change the Cuvette Sample Holder Open the lid of the Sample Holder and locate the Sample Holder Locking Screw Use the Allen Wrench included to...

Страница 6: ...sample cuvette or test tube 4 If you are to test the same sample at other wavelengths repeat step 3 to B 1 4 for each wavelength 5 For each new sample you analyze repeat step 2 to C 4 Output and Data...

Страница 7: ...llowing methods to check wavelength calibration Please note that this test requires the use of a didymium filter or a holmium oxide filter Contact Carolina for availability and pricing of these filter...

Страница 8: ...uld be within 805 809 nm The wavelength accuracy of the unit is 2 nm 9 If a middle wavelength check is desired set the wavelength to 522 nm optional 10 Make sure the cuvette adapter is empty and place...

Страница 9: ...rect reading for the stray light value Troubleshooting PROBLEM Possible Cause Solution Instrument Inoperative Power indicator has no light Power cord not connected to outlet Plug instrument in Dead Po...

Страница 10: ...d properly Check if lamp has been properly installed or has moved during transit Refer to lamp Replacement instructions in this manual Lamp old or defective Replace with a new lamp Sample Holder Misal...

Отзывы: