User’s Manual UTS1000B/T Series
Instruments.uni-trend.com
72 / 76
Fundamental: Set the frequency of the measured carrier signal.
User can change the frequency of basic harmonic by using the numeric key, rotary knob, direction key, or
touch the panel menu. By default, other harmonic measurements will be test at multiples of the specified
fundamental frequency.
Harmonics: Set the harmonic time of the measured carrier to calculate the total harmonic. User can
change the harmonic time by using the numeric key, rotary knob, direction key, or touch the panel menu.
Dwell Time: Sets the dwell time for the specified harmonic. In zero-span, the auto mode strikes a balance
between measurement speed and accuracy; increasing the dwell time will reduce measurement speed
while increasing accuracy. User can change the dwell time by using the numeric key, rotary knob, direction
key, or touch the panel menu.
Meas Preset: Reset all parameters of the current measuring mode to factory setting.
6. System Prompt and Troubleshooting
System Prompt
All the operating mode parameter setting are invalid, use the default value.
Name
Minimum
Maximum
Center Frequency
50Hz
3.209999950GHz
Span
Zero span
Full span
Start Freq
0Hz
3.209999900GHz
Stop Freq
100Hz
3.21GHz
Freq Offset
-100GHz
100GHz
CF Step
1Hz
3.21GHz
Ref Level
-100 dBm
30 dBm
Attenuation
0 dB
51 dB
Scale/Div
0.1dB
20dB
Ref Level Offset
-327.6dB
327.6dB
RBW
1Hz
1MHz
VBW
1Hz
1MHz
VBW:3dB RBW
0.000001
1000000
Sweep Time
1 ms
4 ks
Points
1
10001
Trace Operation Offset
-100dB
100dB
Mark Frequency
0Hz
3.21GHz
Mark
△
Frequency
-3.21GHz
3.21GHz