User’s Manual UTS1000B/T Series
Instruments.uni-trend.com
53 / 76
All Markers Off: Close all the marker point.
Marker->: Use the value of the current cursor to set other system parameters of the spectrum analyzer
(such as center frequency, reference level, etc.). If no cursor is available at the current, press Marker menu
to activate a cursor automatically.
1. ->CF: Set the center frequency of the spectrum analyzer to the frequency of the current cursor.
When “Normal” cursor is selected, the center frequency is set to the frequency of the cursor.
When “Difference” cursor is selected, the center frequency is set to the frequency of the difference
cursor.
In zero span mode, this function is not work.
2. ->CF Step: Set the stepped center frequency of the spectrum analyzer to the frequency of the current
cursor.
When “Normal” cursor is selected, the stepped center frequency is set to the frequency of the cursor.
When “Difference” cursor is selected, the stepped center frequency is set to the frequency difference
between the difference cursor and the reference cursor.
In zero span mode, this function is not work.
3. ->Start: Set the start frequency of the spectrum analyzer to the frequency of the current cursor.
When “Normal” cursor is selected, the start frequency is set to the frequency of the cursor.
When “Difference” cursor is selected, the start frequency is set to the frequency of the difference
cursor.
In zero span mode, this function is not work.
4. ->Stop: Set the stop frequency of the spectrum analyzer to the frequency of the cursor.
When “Normal” cursor is selected, the stop frequency is set to the frequency of the current cursor.
When “Difference” cursor is selected, the stop frequency is set to the frequency of the difference
cursor.
In zero span mode, this function is not work.
5. ->Ref Lvl: Set the reference level of the spectrum analyzer to the amplitude of the valid marker, and
move the marker point to the reference level (top of the grid).
When “Normal” cursor is selected, set the marker amplitude of the spectrum analyzer to the amplitude
of the current reference level.
When “Difference” cursor is selected, set reference level to the amplitude difference between
markers.
Marker Table: Turn on/off marker Table
When the marker Table is opened, all open cursors are displayed as a Table in the window of the split
screen. The display includes cursor number, marker mode, marked trace number, X-axis scale type, X-axis
reading and amplitude. The cursor table can be used to view the measurement values of multiple
measurement points.