UTD1000 User Manual
Appendix A : Performance Indicators
Unless otherwise specified, all technical specifications apply to probes with an attenuation setting of 10X and the
UTD1000 Series Digital Oscilloscopes.
To verify that your Oscilloscope meets specifications, it must first meet the following conditions :
Ƶ
The Oscilloscope must have been operating continuously for thirty minutes within the specified operating
temperature.
Ƶ
I
f the operating temperature changes by more than 5°C, you must perform the “Self Cal” operation, accessible
through the System Functions menu. All specifications are guaranteed unless noted “Typical”.
Technical Indicators :
Chapter 6 Appendixes
Sampling modes
Sampling rates
Real time
25
0
MS/s (UTD1025C
/
UTD1042C/UTD1062C)
500MS/s(UTD1082C/UTD1102C)
1GS/s (UTD1152C/UTD1202C)
Sampling, peak detect, average
Equivalent
25GS/s
When all channels have made N samplings simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 to 256.
Sampling Modes
Sampling
Average