UTD1000 User Manual
Definitions :
Display Type :
Vector display fills the space between
adjacent sample points in the display. Dot display shows
only the sampling points.
Refresh rate : Refresh rate is the number of times the
digital storage Oscilloscope refreshes the waveform
display per second. The refreshing speed affects the
capability to observe signal movements speedily.
X-Y Mode
This mode can be activated only when Channel A and
Channel B are both on. After selecting the X-Y display
mode, the horizontal axis will display the A signal, while
the vertical axis will display the B signal.
Caution :
In the normal X-Y mode, you can adjust waveform
amplitude with VOLTS/DIV of the two channels.
Adjusting the s/DIV time base will result in a lissajous
figure of better display quality.
The following functions are disabled in the X-Y display
mode :
Ƶ
Automatic measurement mode
Ƶ
Cursor measurement mode
Ƶ
Reference or maths waveform
Ƶ
Vector display type
Ƶ
Trigger control
Definitions :
Y-T Mode :
In this mode the Y axis indicates voltage
and the X axis indicates time.
X-Y Mode :
In this mode the X axis indicates Channel A
voltage and the Y axis indicates Channel B voltage.
Slow Scanning Mode :
When horizontal time base
control is set at 50ms/div or slower, the unit will operate
in the slow scan sampling mode. When observing low
frequency signals in slow scanning mode, it is advised
to set channel coupling to DC.
S/ns :
A horizontal scaling (time base) unit. If waveform
sampling is stopped (by pressing the [RUN/STOP]
button), time base control can expand or compress the
waveform.