User’s Manual UT5583 Insulation Resistance Tester
Instruments.uni-trend.com
44
/
47
13.3.2 Set IP Address
Move cursor to [IP Address] field, use function key to select;
Input box will be pop out, use numeric keyboard or use rotary knob and left and right cursor to adjust
IP address.
And then press [OK] key to confirm the change; press [Esc] key to cancle the change.
14. Specification
14.1 Technology Index
Measurement parameter
Insulation resistance, leakage current
Output voltage
Range control 1V
~
1000V (DC):
When voltage ≥ 10V, adjusting step 1V
When voltage < 10V, adjusting step 0.1V
Accuracy of voltage
When voltage ≥ 10V, 1% ± 1V
When voltage < 10V, 10% ± 0.1V
Measuring range
Resistance: 10 kΩ
~
10 TΩ
Current: 100 pA
~
250 uA
Accuracy of measurement
1V ≤ Voltage
< 10V
5% (less than 1M); 1% (1M~100M); 5% (100M~10G);
10% (greater than 10G)
10V ≤ Voltage
≤1000V
5% (less than 1M); 1% (1M~1G); 3% (1G~10G);
5% (10G~100G); 10% (100G~1T); 15% (greater than 10G)
Test speed
Fast speed: 30 ms/time; Middle speed: 100 ms/time; Slow speed: 500
ms/time
Range mode
Automatic, lock, nominal (select the best range according to the
comparator setting)
Maximum of charging current 25 mA ± 5 mA