User’s Manual                                                                                                                                                        UT5583 Insulation Resistance Tester 

Instruments.uni-trend.com                                                                                                                                                                                   

 

33

 / 

47

 

 

13 

TRIG 

Test trigger input terminal   

Edge signal, trigger edge tyoe can select in <Setup> page, the default 
is falling edge. 
Low level maintains 1ms.

   

 
Table 12-3 Pin Definition of Voltage Terminal in Single Comparison Mode 

Pin 

Name 

Description 

19 

ISO-COM 

Common ground, not allow to float. 
Be sure to be reliably connected to the ground (COM) port of the power 
supply of an external controller (such as a PLC). 

20 
21 
24 

Internal 
ISO-VCC 
Output 

Positive output of internal VCC power supply 
Internal isolation power voltage: 5V, 0.2A, 1Wmax 
Do not connect if it is not necessary. 

25 

 

 

1. Pin P24/P25 is output terminal of internal isolation power. 
2. Warning: Set pin P24/P25 to be floating when normal connecting to PLC. 
3. Internal power voltage is limited (5V

0.2A, 1Wmax), it cannot drive the power 

relay or high power LED. 
4.  Pin  P24/P25  has  built-in  automatic  self-recovery  fuse,  it  will  cut  off  the 

connecting of ISO-VCC when it excceds 0.5 A, long short circuit will cause the 
fuse to burn out. 

 

12.2.2 Time Sequence of Single Comparison Mode 

Figure 12-1 Time Squence of Start/Stop in Single Comparative Mode   

 

 

Содержание UT5583

Страница 1: ...User s Manual UT5583 Insulation Resistance Tester...

Страница 2: ...form at the same specifications as brand new products All replacement parts modules and products become the property of UNI T The customer refers to the individual or entity that is declared in the gu...

Страница 3: ...for merchant ability or applicability purposes For violation of this guarantee regardless of whether UNI T and its distributors are informed that any indirect special incidental or consequential damag...

Страница 4: ...e is designed for professional users and responsible organizations for measurement purposes Do not use this device in any way not specified by the manufacturer This device is only for indoor use unles...

Страница 5: ...t before you proceed to operation Note It indicates potential problems which may cause failure of this device if you fail to follow a certain procedure or condition If the Note sign is present all con...

Страница 6: ...rge buildings machine tools and power distribution boards at industrial sites workshops CAT IV Three phase public power unit and outdoor power supply line equipment Equipment designed to initial conne...

Страница 7: ...y Please use the AC power supply specified for this device Please use the power cord approved by your country and confirm that the insulation layer is not damaged Electrostatic prevention This device...

Страница 8: ...sonnel of UNI T for testing Any maintenance adjustment or parts replacement must be done by the relevant personnel of UNI T Cooling Do not block the ventilation holes at the side and back of this devi...

Страница 9: ...V adjusting step 1V When voltage 10V adjusting step 0 1V Accuracy of voltage When voltage 10V 1 1V When voltage 10V 10 0 1V Measurement range Resistance 10 k 10 T Current 100 pA 250 uA Accuracy of mea...

Страница 10: ...cuit zero clearing Contact detection for capacitive materials Digit display function 5 digits 4 digits Comparator function sorting qualification and disqualification Sound setting of sorting judgement...

Страница 11: ...nal trigger signal lock key signal input port of built in pull up resistor Output all sorting comparison result signals test completion signal EOM large current drive output can direct drive relay RS...

Страница 12: ...minal is used to connect test cable for measurement positive terminal current sampling terminal negative terminal voltage output terminal dangerous high voltage GND Ground terminal block the object to...

Страница 13: ...Figure 4 2 Rear Panel No Description 1 Protective ground terminal 2 Fuse 3 Power socket 4 AC 220 110V adaptor 5 Handler interface 6 RS 232C RS 485 interface option 7 LAN interface 8 Footswitch interf...

Страница 14: ...annot connect to positive and negative 4 3 3 Connecting UNT with Polarity UNT unit under test with polarity such as the capacitor with positve and negative polarity It must be meaured as the following...

Страница 15: ...tee the accuracyandstability ofmeasurement please ensure that the temperature and humidity of the environment meet the required conditions 4 4 Open Circuit Zero Clearing It is used to perform open cir...

Страница 16: ...Pluggable wiring terminal 1 piece Footswitch Download manual 1 piece User s manual 0 piece Electronic user s manual can download from the official website 5 2 Power Requirements The instrument is dei...

Страница 17: ...Do not clean the internal of the instrument Note Do not use solvent alcohol or gasoline to clean instrument 5 5 Handle Handle is adjustable and can adjust to four positions hold two sides of the hand...

Страница 18: ...age Test display page is mainly used to display measured results test state and sorting results 5 common function can be set at this page including Voltage output voltage setting Speed test speed sett...

Страница 19: ...equipment online test please choose middle or fast speed In general use slow speed when test device Setting Steps of Test Speed 1 Press Test key to enter the test page or press Setup key to enter the...

Страница 20: ...function key at the bottom of the screen to select Range Function Key Setting Automatic Range The instrument will automatically select the range Locked Range The instrument will be locked at the curr...

Страница 21: ...LFAIL Open circuit poor contact OPEN 6 Insulation resistance display It displays in white character during the test and in gray character during discharge stop state Note The instrument will direct di...

Страница 22: ...ument will automatically turn to the test state If Charge time sets to OFF press On Off key to direct enter the test state Test state the state displays test and the test indicator is illuminated with...

Страница 23: ...ady 6 The network is connected 7 Sorting sound and key sound is enabled Sorting sound and key sound is disabled Only key sound is enabled Only sorting sound is enabled 8 Time display 6 3 Screenshot Th...

Страница 24: ...and OK key to select the parameter Table 7 1 Description of Setup Page Item Setting Default Setting Description Test Time Continuous 0 5s 2s 10s 30s 60s Continuous In continous the instrument enters c...

Страница 25: ...n receive an edge pulse the detail refer to Trigger Edge Setting from Handler interface on the rear panel The instrument will perform a measurement after receive signal command and stay in the wait st...

Страница 26: ...esistance can be set when the comparator is enabled Use numeric keyboard and OK key to input Upper Limit of Resistance The upper limit of resistance can be set when the comparator is enabled Use numer...

Страница 27: ...e Instrument uses 24 hour system change the date and time by function keys Volume Low middle high Low Beeper s volume Sound Key ON OFF ON The switch of key sound Backlight Brightness 10 30 50 70 90 10...

Страница 28: ...est is completed without the need for the host to send the FETCH command IP Address 192 168 030 036 It can set by numeric keyboard it will be used when select LAN interface Port Number 502 It is defau...

Страница 29: ...USB memorizer User can quickly set batch setting for the instrument and support save the setup parameter to external USB Other instrument can read the test setup parameter from USB AUTO LOAD File 1 bo...

Страница 30: ...quirements refere to Screenshot function After the test is finished the instrument will automatically save the test data to USB 11 1 USB Data Storage Mechanism The instrument only supports set compara...

Страница 31: ...g output PASS FAIL EOM test completion signal and TRIG external trigger initiation input Depending on the comparator mode single and period the instrument provides two sets of handler interface test s...

Страница 32: ...d 1 HI Over the upper limit disqualified output of comparator 0 HI Low level is valid 2 LO Over the lower limit disqualified output of comparator 0 LO Low level is valid 16 OPEN Poor contact open circ...

Страница 33: ...21 24 Internal ISO VCC Output Positive output of internal VCC power supply Internal isolation power voltage 5V 0 2A 1Wmax Do not connect if it is not necessary 25 1 Pin P24 P25 is output terminal of...

Страница 34: ...test state 30ms T9 Stop signal valid and low level holding time 30ms T1 Trigger pulse width 1ms T2 Measurement period Trigger delay Relate to Setup refer to section 7 Trigger Delay T3 AD conversion t...

Страница 35: ...when sampling and turn to low leven when sampling is finished 14 PASS Qualified output of comparator 0 PASS Low level is valid 15 FAIL Total disqualified output of comparator 0 FAIL Low level is vali...

Страница 36: ...20 21 24 Internal ISO VCC Output Positive output of internal VCC power supply Internal isolation power voltage 5V 0 2A 1Wmax Do not connect if it is not necessary 25 1 Pin P24 P25 is output terminal...

Страница 37: ...after testing 12 4 Wiring Method Power Supply Built in isloation power supply external power is no need but common ground ISO COM is a must ISO COM P19 P21 Output Signal Optoelectronic isolation with...

Страница 38: ...ents uni trend com 38 47 12 4 2 Schematic Diagram of Output Terminal Figure 12 5 Schematic Diagram of Output Terminal 12 4 3 Connection of Input Circuit Figure 12 2 Connect to Switch Figure 12 7 Relay...

Страница 39: ...istance Tester Instruments uni trend com 39 47 12 4 4 Connection of Output Circuit Figure 12 9 Control Relay The maximum of EX VCC is 30V Figure 12 10 Control Luminous Diode or Optoelectronic Isolator...

Страница 40: ...LC Negative Common Terminal 12 5 External Connecting Footswitch Figure 12 14 Wiring Terminal of Footswitch Installation Method Connect the switch between the two terminals of FOOT C Instructions Compa...

Страница 41: ...on the RS 232 standard a 25 core connector is used in each port today s computers basically use a 9 core connector The common RS 232 signal as shown in the following table Table 13 1 Common RS 232 Si...

Страница 42: ...nfiguration page Data bit 8 bits Stop bit 1 bit Parity bit no 13 2 RS 485 Interface Setting The option RS485 interface of instrument and support ModBus RTU protocol The standard interface of instrumen...

Страница 43: ...in Function 8 A 9 B 13 3 LAN Interface Setting Figure 13 3 LAN Connector on the Rear Panel Connect LAN cable to the LAN connector of the instrument Green LED illumed connecting blinking communicating...

Страница 44: ...current Output voltage Range control 1V 1000V DC When voltage 10V adjusting step 1V When voltage 10V adjusting step 0 1V Accuracy of voltage When voltage 10V 1 1V When voltage 10V 10 0 1V Measuring ra...

Страница 45: ...record Control interface HANDLER FOOT Communication protocol SCPI Modbus RTU Communication interface RS232C USB HOST LAN option RS485 Weight 3 32 kg Size 382 3 215 88mm 14 2 Accuracy of Resistance Te...

Страница 46: ...requently according to the operating condition Follow these steps to clean the external surface of the instrument a Please use a soft cloth to wipe the dust outside the instrument b When cleaning the...

Страница 47: ...quential loss 15 3 Appendix C Contact Us If the use of this product has caused any inconvenience if you in mainland China you can contact UNI T company directly Service support 8am to 5 30pm UTC 8 Mon...

Отзывы: