7. Test <Setup> Page
In the stop state, press [Setup] key to enter setup page;
In <Setup> page, the instrument is not perform test.
Figure 7-1 <Test> Setup Page
The setting of output voltage, test speed and measuring range refers to section 6.
Parameter setting: press [Setup] key to enter <setup> page, use arrow key to move the cursor to the
setting item, press the function key at the bottom of the screen or us numeric keyboard and [OK] key to
select the parameter.
Table 7-1 Description of <Setup> Page
Item
Setting
Default Setting
Description
Test Time
Continuous, 0.5s, 2s,
10s, 30s, 60s
Continuous
In continous, the instrument enters
continuous test state.
Charge Time
OFF, 2s, 5s, 10s, 30s,
60s
OFF
Or use numeric keyboard to input arbitrary
time.
Discharge
Time
OFF, 2s, 5s, 10s, 30s,
60s
OFF
Or use numeric keyboard to input arbitrary
time.
Display Mode
Resistance, resistance
and current
Resistance
Resistance ---- only display insulation
resistance value in test page.
Resistance and current ---- only display
insulation resistance and leakage current
value in test page.
Contact
Inspection
ON, OFF
OFF
Whether perform the contact inspection or
not