background image

5

6

1.2 Features

1.3.1 Insulation Resistance Specifications

Rated 
voltage

UT512D
measurement range

UT512E
measurement range

Accuracy

Over-range 
indication

Short-circuit
current

1. Insulation resistance range up to 1.0TΩ (UT512D) and 2.5TΩ (UT512E)
2. Output rated voltage ranges including 250V, 500V, 1000V and 2500V
3. Short-circuit current: About 3.5mA (UT512D) and 5.0mA (UT512E)
4. Insulation resistance testing (IR)
5. AC/DC voltage testing (ACV, DCV)
6. Capacitance testing (CAP) (UT512E)
7. Low resistance testing  ( R ) (UT512E)
8. With stepping set at 10% of rated voltage for each range.
9. Polarization Index (PI) and Dielectric Absorption Ratio (DAR)
10. Resistance comparison function (COMP) with upper and lower resistance 
      limit setting and over-range indication.
11. Timer testing mode
12. Automatic current display
13. External voltage detection function to monitor live voltage of measured object.
14. Timer function to record testing time automatically.
15. Automatic discharge and high voltage alarming functions.
16. Analog bar graph to display tested insulation resistance
17. Manual and automatic power off functions
18. The ability to delete and store 999 groups of testing data.
19. Data upload function to enable uploading data to computer via USB cable for 
      data analysis.
20. LCD-backlit function
21. 5.1-inch LCD display
22. 14.8V/5200mAh rechargeable lithium battery pack (UT512E); 8 PCS LR14 
      alkaline batteries (UT512D)

Error limit: ± (a% of r b digits), one-year warranty

Environment temperature: 23±5

Environment humidity: 45~75%RH, In the table below: The humidity must be less 
than 50%RH when testing insulation objects with a resistance greater than 50GΩ 
in the parameters. 
Temperature coefficient: Test outside the temperature range of the indicator (i.e. 
above 28 degrees or below 18 degrees), with an additional testing error of+/-0.25% 
for each degree Celsius.

1.3 Technical Specifications

250V

500V

1000V

2500V

>

<

0.25MΩ

0.25MΩ~4.99GΩ

5.00GΩ~24.9GΩ

25.0GΩ~100GΩ

<0.50MΩ

0.50MΩ~9.99GΩ

10.0GΩ~49.9GΩ

50.0GΩ~200GΩ

<1.00MΩ

1.00MΩ~19.9GΩ

20.0GΩ~99.9GΩ

100GΩ~400GΩ

<2.50MΩ

2.50MΩ~49.9GΩ

50.0GΩ~249GΩ

250GΩ~1.00TΩ

<0.25MΩ

0.25MΩ~4.99GΩ

5.00GΩ~49.9GΩ

50.0GΩ~250GΩ

<0.50MΩ

0.50MΩ~9.99GΩ

10.0GΩ~99.9GΩ

100GΩ~500GΩ

<1.00MΩ

1.00MΩ~19.9GΩ

20.0GΩ~199GΩ

200GΩ~1.00TΩ

<2.50MΩ

2.50MΩ~49.9GΩ

50.0GΩ~499GΩ

500GΩ~2.50TΩ

±(

5%+5)

±(20%+10)

±(

5%+5)

±(20%+10)

±(

5%+5)

±(20%+10)

±(

5%+5)

±(20%+10)

For reference only

For reference only

For reference only

For reference only

For reference only

For reference only

For reference only

For reference only

UT512D: 
About 3.5mA

UT512E: 
About 5.0mA
current 
with load:
1mA~1.2mA
(250V,
0.25MΩ;
500V,
0.5MΩ;
1000V,
1.0MΩ;
2500V,
2.5MΩ;

1 TΩ (Tera ohm) =1000GΩ=10¹²Ω

1 GΩ (Giga ohm) =1000MΩ=10⁹Ω

1 MΩ (Mega ohm) =1000KΩ=10⁶Ω

Note: For insulation resistance measurement, if the measured capacitive reactance 

is greater than about 100nF, it may cause reading to fluctuate significantly.

"Overrange warning: For example, in the 1000V voltage output range of UT512D, 

if the range is exceeded, it will display >400GΩ.".

UT512D/E User Manual

UT512D/E User Manual

Содержание UT512D

Страница 1: ...UT512D UT512E 2 5kV Insulation Resistance Tester User Manual P N 110401111817X...

Страница 2: ...Trend guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misu...

Страница 3: ...ainly used to measure insulation resistance of cable motor power generator transformer mutual inductor high voltage switch lightning arrester and more it is an ideal measurement meter for electric pow...

Страница 4: ...T512E 8 PCS LR14 alkaline batteries UT512D Error limit a of reading b digits one year warranty Environment temperature 23 5 Environment humidity 45 75 RH In the table below The humidity must be less t...

Страница 5: ...ications UT512E Function Measurement range Accuracy Remark Low resistance measurement Open circuit voltage 5V Note If the measured resistance is 20 the buzzer alarms if 660 660 is displayed The short...

Страница 6: ...bidden to use the tester without battery cover set in place otherwise it may pose a risk of electric shock Keep fingers behind the finger guard when perform measurement do not touch exposed wire conne...

Страница 7: ...plug red lead GUARD Grounding protection outlet single plug green lead EARTH High resistance measurement outlet single plug black lead LCD segmented display screen Functional buttons USB port Carryin...

Страница 8: ...testing Bluetooth Insulation resistance measurement mode AC DC voltage measurement mode Power on off Increase select high range or previous group of data Decrease adjust time and resistance cyclically...

Страница 9: ...ress this button to switch to insulation resistance testing function V UT512E AC DC voltage testing button Without high voltage output short press V button to switch to AC DC voltage testing mode The...

Страница 10: ...measurement Data deletion button In READ state short press CLR and then press ENTER to delete current data press ESC to exit data deletion function In READ state the symbol CLR and the storage capaci...

Страница 11: ...rged if low battery occurs please replace with battery in same type 9 2 Battery Charging UT512E When charging with battery charger in power on state the battery power indicator and the charging symbol...

Страница 12: ...hange Press TEST to perform test the LCD shows basic elements including battery power high voltage warning symbol leakage current real time output voltage measured insulation resistance testing value...

Страница 13: ...tion resistance in Time1 9 3 1 4 Dielectric Absorption Ratio Measurement Dielectric Absorption Ratio DAR refers to the value of insulation resistance in 1 minute to that in 15 seconds It takes 1 minut...

Страница 14: ...s if insulation resistance is less than comparison resistance the symbol FAIL will appear on the LCD otherwise PASS appears To return to the continuous measurement interface please press COMP in COMP...

Страница 15: ...ltage input the danger symbol flashes in such case do not press the test button to measure resistance as shown in figure 9 3 4 2 Under resistance measurement function connect the black grounding lead...

Страница 16: ...rminal with the exterior conductor of inner insulation layer and the unused cable Surface leakage current is conducted to the safety terminal so as to eliminate the surface leakage current at the meas...

Страница 17: ...Red XI Maintenance Clean the casing 1 Wipe the surface using soft cloth or sponge dipped with water 2 Do not immerse the tester in water to avoid tester damage 3 Do not store the tester if it is wet 4...

Отзывы: