UT3500 series
User Manual
39
2. Use the cursor keys to move the cursor to the [R-COMP] or [V-COMP] field to select
ON;
3. Move the cursor to the [MODE] field, and use the function keys at the bottom of the
screen to select the desired comparison mode among the three according to your
needs.
7.1.3 [Nominal Value] Input
The absolute value and percentage comparison mode must enter the nominal value. As
for direct reading value comparison mode, the nominal value does not participate in
calculation.
Setting Steps:
1. Press the [Test] or [Setup] key, and select [COMP SET] with the function key at the
bottom of the screen to enter the <COMP SET> page;
2. Use the cursor keys to move the cursor to the [R-COMP] or [V-COMP] field to select
ON;
3. Move the cursor to the [MODE] field, select the absolute value or percentage
comparison method, use the number keys to enter the data, and the unit should use
the function keys to select.
7.1.4 [Lower] and [Upper] Setting
Setting Steps:
1. Press the [Test] or [Setup] key, and select [COMP SET] with the function key at the
bottom of the screen to enter the < COMP SET> page;
2. Use the cursor keys to move the cursor to the [R-COMP] or [V-COMP] field to select
ON;
3. Move the cursor to the corresponding [UPPER] or [LOWER] field, use the number keys
to directly enter the data, and use the function keys to select the unit;
4. Note: The percentage
△
% mode does not need to select the unit magnification,
please enter the percentage value; for the absolute value
△
and the direct reading
value SEQ mode, please use the function keys to select the unit.
The instrument shares the same storage space for the three comparison modes, so
the comparator data needs to be reset after switching the comparison mode.
7.2 Display and Discrimination
7.2.1 Display after Opening the Comparator
When the comparator setting is completed, press [Test] to enter the main page of <TEST>,
as shown in Figure 7-2-1.