7.1 Comparator Setting
Figure 7-1 <COMP SET> Page
7.1.1 [Beep] Setting
The beep function is effective only after the comparator function is turned on.
There are three types of beep settings: PASS, FAIL, OFF.
PASS: the buzzer will sound when the sorting result is qualified. FAIL: the buzzer will sound
when the sorting result is unqualified.
Setting Steps:
1. Press the [Test] or [Setup] key, and select [COMP SET] with the function keys at the
bottom of the screen to enter <COMP SET> page;
2. Use the cursor keys to move the cursor to the [BEEP] field;
3. According to your own needs, use the function keys at the bottom of the screen to
select PASS, FAIL or OFF.
7.1.2 Comparison Mode Selection
The instrument can compare resistance and voltage simultaneously or separately. There
are 3 comparison modes to choose from:
a) Absolute Value Comparison ABS (
△
): Absolute Value
△
= Measured Value-Nominal
Value
b) Percentage Comparison PER (
△
%): Percentage
△
% = (Measured Value-Nominal
Value) / Nominal Value
×
100 %
c) Direct-Reading Comparison (SEQ): The direct-reading value SEQ comparison uses the
direct-reading measurement value to compare with the upper and lower limit ranges, so
the nominal value is not required to participate in the calculation.
Setting Steps:
1. Press the [Test] or [Setup] key, and select [COMP SET] with the function keys at the
bottom of the screen to enter <COMP SET> page;