background image

23

24

Main display: Standby mode prompt
Auxiliary display 1: Active power
Auxiliary display 2: Time

Main display: Measured electric energy
Auxiliary display 1: Active power
Auxiliary display 2: Elapsed time

Main display: Measured electric energy
Auxiliary display 1: Active power
Auxiliary display 2: Elapsed time

Start accumulating

Restart

Short press

Start accumulating

Short press

Short press

Measurement interface of compare function of electric meter

Short press

Restart

Start accumulating

Start accumulating

Short press

Short press

Main display: Standby mode prompt
Auxiliary display 1: Active power
Auxiliary display 2: Time

Main display: Operation prompt
Auxiliary display 1: Active power
Auxiliary display 2: Time
Auxiliary display 3: Theoretical value

Main display: Actual measured 
value—theoretical value
Auxiliary display 1: Actual measured value, 
theoretical value
Auxiliary display 2: Actual measured value
Auxiliary display 3: Theoretical value

  

Operate the compare function of electric meter

     For comparing mechanical meter: Press “Hold” to start accumulation when the round disk 
     rotates for one round; press “HOLD” again to stop accumulation when the round disk rotates 
     for another round, as shown below:

Mechanical Meter

Round disc rotates for one round

Round disc

Start accumulating

Stop accumulating

UT219P User Manual

UT219P User Manual

For comparing electronic meter: Press “Hold” to start accumulation when the LED flashes for 
one time, press “HOLD” again to stop accumulation when the LED flashes again, as shown 
below:

Electronic Meter

LED flashes for one time

Start accumulating

Stop accumulating

For comparing fixed electric energy meter: Take 0.1kWh as an example, press “HOLD” to start 
accumulation when the value changes by 0.1kWh; press “HOLD” again to stop accumulation 
when the value changes by 0.1kWh again, as shown below:

Decimal place: 0.1 kWh

Start accumulating

Stop accumulating

Note:

   Phase angle is in reference to voltage.
   Do not input voltage over 1000V. It is possible to measure higher voltage, but it may damage 
   the clamp.
   Avoid electric shock when working with high voltage.
   Disconnect test leads with the measured circuit after all measurement operations are 
   completed.
   The high voltage warning symbol “     ” will appear on the LCD if the measured voltage is 
   over 30V AC; the red backlight flashes if the measured voltage is over 1000VAC.
   Do not release the trigger abruptly when pressing and holding it. Hall sensor is sensitive to 
   magnet, heat and mechanic stress, impact can cause reading variation in a short time.
   To ensure measurement accuracy, please place the measured conductor at the center of the 
   clamp jaws. Error of ±1.0% of reading will be produced if the measured conductor is not 
   placed at the center of the clamp jaws.
   Incorrect wire connection will cause negative active power and cause the red backlight to 
   flash, please check if the connections of voltage input terminal and clamp jaws are correct.

Содержание UT219P

Страница 1: ...P N 110401111507X UT219P AC Digital Power Clamp Meter User Manual...

Страница 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Страница 3: ...1000V AC 4 1000V AC RMS voltage measurement 15Hz 1000Hz frequency response range 5 1000A AC RMS current measurement 15Hz 1000Hz frequency response range 6 Ability to measure and display RMS and peak...

Страница 4: ...uracy Remove the battery if the product is not used for a long time 10 Do not alter the internal wiring to avoid product damage and safety hazard 11 Do not keep or use the product in environments with...

Страница 5: ...ctor 4 3P3W 3 phase 3 wire measurement mode 5 3P4W 3 phase 4 wire measurement mode 6 UNBALANCED Unbalanced measurement mode 7 MAX Maximum value measurement 8 MIN Minimum value measurement 9 AVG Averag...

Страница 6: ...ximum value minimum value average value on the main display and to record the time of capturing maximum and minimum values Long press this button to exit maximum minimum value measurement The measurem...

Страница 7: ...ip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to A V press SELECT to select AC voltage measurement and connect the alligator clips w...

Страница 8: ...lamp jaws are correct L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase 3 wire power measurement 3P3W balance Load 4 3 phase power measurement 4 1 3 phase 3 wire 3P3W balanced mode Fig 8 Fig 8 UT...

Страница 9: ...voltage Auxiliary display 3 AC current Short press Short press Short press Short press Short press 4 3 3 phase 3 wire 3P3W unbalance mode Fig 10 L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase...

Страница 10: ...alligator clip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to 3PW press to select 3P4W unbalance mode then the clamp enters P1 wire c...

Страница 11: ...ower Auxiliary display 1 P1 prompt Auxiliary display 2 Voltage Auxiliary display 3 Current Main display Second active power Auxiliary display 1 P2 prompt Auxiliary display 2 Voltage Auxiliary display...

Страница 12: ...at the center of the clamp jaws Error of 1 0 of reading will be produced if the measured conductor is not placed at the center of the clamp jaws If zero crossing of harmonic does not meet the zero cr...

Страница 13: ...tronic meter Press Hold to start accumulation when the LED flashes for one time press HOLD again to stop accumulation when the LED flashes again as shown below Electronic Meter LED flashes for one tim...

Страница 14: ...ace of wire connection Fig 13 within 10 seconds connect test lead from V terminal with L3 live wire and connect test lead from COM terminal with L2 live wire then three kinds of testing results will b...

Страница 15: ...al error of temperature coefficient is 0 1 specified accuracy C To ensure measurement accuracy please place the measured conductor at the center of the clamp jaws Error of 1 0 of reading will be produ...

Страница 16: ...angle error 3 5 Phase angle 15Hz 40Hz 70Hz 1000Hz 40Hz 70Hz 1000A AC 1000Vrms 0 1 3 5 5 Range Resolution Accuracy Overload protection Remark Min measured voltage 30V Max measured current 10A 180 adva...

Страница 17: ...ment result viewing button control and other operations can be achieved through UNI T Smart Measurement APP 3 2 UNI T Smart Measurement APP has multiple functions including Bluetooth communication dat...

Отзывы: