7. Phase sequence detection (Fig. 13)

L1

/

R

/

A

L2

/

S

/

B

L3

/

T

/

C

Black

Red

Power supply

Phase detection

Load

Black

Red

1) Connect red test lead (connected with red alligator clip already) with “V” terminal; and black 
    (connected with black alligator clip already) with “COM”.
2) Set the rotary switch to “Phase Detect” to enter the first prompt interface of wire connection, 
    as shown below:

“L2” on bottom left corner indicates test lead from COM terminal connects with L2 live wire, 
“L1” bottom right corner indicates test lead from V terminal connects with L1 live wire.

3) According to the first prompt interface of wire connection (Fig. 13), connect test lead from V 
    terminal with L1 live wire, then connect test lead from COM terminal with L2 live wire. When 
    the voltage is locked, the clamp enters the second prompt interface of wire connection, as 
    shown below:

“L2” on bottom left corner indicates test lead from COM terminal connects with L2 live 
wire, “L3” bottom right corner indicates test lead from V terminal connects with L3 live wire, 
then the prompt time “10s” on top right corner starts to decrease.

UT219P User Manual

UT219P User Manual

Fig. 13

4) According to the second prompt interface of wire connection (Fig. 13), within 10 seconds, 
    connect test lead from V terminal with L3 live wire, and connect test lead from COM terminal 
    with L2 live wire, then three kinds of testing results will be displayed.
    For positive sequence, “L1”, “L2” and “L3” light up clockwise, as shown below:

Note:

    Do not input voltage over 1000Vrms. It is possible to measure higher voltage, but it may 
    damage the clamp.
    Avoid electric shock when working with high voltage.
    Disconnect test leads with the measured circuit after all measurement operations are 
    completed.
    The high voltage warning symbol “    ” will appear on the LCD if the measured voltage is over 
    30V AC; the red backlight flashes if the measured voltage is over 1000VAC.

5) Press “HOLD” to return to initial interface to start phase sequence detection.

8. Other functions

 

Auto power off:

    

The clamp powers off automatically after 15 minutes of inactivity. To awaken the clamp, please 
press SELECT button in Auto-off state. To disable the Auto-off function, please hold down 
SELECT button and power on the clamp. Restart the clamp to enable the Auto-off function.

    Battery voltage detection:
When the battery voltage is:
>3.6V, “        ” appears.
3.4V~3.6V, “

” appears.

        

3.2V~3.4V, “

” appears.

        

3.0V~3.2V, “

” appears.

        

2.8V~3.0V, “

” appears.

        

<2.8V, the clamp powers off automatically.

    For inverse sequence, "L1", "L2" and "L3" light up counterclockwise, and red backlight appears, 
    as shown below:

Содержание UT219P

Страница 1: ...P N 110401111507X UT219P AC Digital Power Clamp Meter User Manual...

Страница 2: ...end guarantees that the product is free from any defect in material and workmanship within one year from the purchase date This warranty does not apply to damages caused by accident negligence misuse...

Страница 3: ...1000V AC 4 1000V AC RMS voltage measurement 15Hz 1000Hz frequency response range 5 1000A AC RMS current measurement 15Hz 1000Hz frequency response range 6 Ability to measure and display RMS and peak...

Страница 4: ...uracy Remove the battery if the product is not used for a long time 10 Do not alter the internal wiring to avoid product damage and safety hazard 11 Do not keep or use the product in environments with...

Страница 5: ...ctor 4 3P3W 3 phase 3 wire measurement mode 5 3P4W 3 phase 4 wire measurement mode 6 UNBALANCED Unbalanced measurement mode 7 MAX Maximum value measurement 8 MIN Minimum value measurement 9 AVG Averag...

Страница 6: ...ximum value minimum value average value on the main display and to record the time of capturing maximum and minimum values Long press this button to exit maximum minimum value measurement The measurem...

Страница 7: ...ip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to A V press SELECT to select AC voltage measurement and connect the alligator clips w...

Страница 8: ...lamp jaws are correct L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase 3 wire power measurement 3P3W balance Load 4 3 phase power measurement 4 1 3 phase 3 wire 3P3W balanced mode Fig 8 Fig 8 UT...

Страница 9: ...voltage Auxiliary display 3 AC current Short press Short press Short press Short press Short press 4 3 3 phase 3 wire 3P3W unbalance mode Fig 10 L1 R A L2 S B L3 T C Red Black Power supply AC 3 phase...

Страница 10: ...alligator clip already with V terminal and black connected with black alligator clip already with COM 2 Set the rotary switch to 3PW press to select 3P4W unbalance mode then the clamp enters P1 wire c...

Страница 11: ...ower Auxiliary display 1 P1 prompt Auxiliary display 2 Voltage Auxiliary display 3 Current Main display Second active power Auxiliary display 1 P2 prompt Auxiliary display 2 Voltage Auxiliary display...

Страница 12: ...at the center of the clamp jaws Error of 1 0 of reading will be produced if the measured conductor is not placed at the center of the clamp jaws If zero crossing of harmonic does not meet the zero cr...

Страница 13: ...tronic meter Press Hold to start accumulation when the LED flashes for one time press HOLD again to stop accumulation when the LED flashes again as shown below Electronic Meter LED flashes for one tim...

Страница 14: ...ace of wire connection Fig 13 within 10 seconds connect test lead from V terminal with L3 live wire and connect test lead from COM terminal with L2 live wire then three kinds of testing results will b...

Страница 15: ...al error of temperature coefficient is 0 1 specified accuracy C To ensure measurement accuracy please place the measured conductor at the center of the clamp jaws Error of 1 0 of reading will be produ...

Страница 16: ...angle error 3 5 Phase angle 15Hz 40Hz 70Hz 1000Hz 40Hz 70Hz 1000A AC 1000Vrms 0 1 3 5 5 Range Resolution Accuracy Overload protection Remark Min measured voltage 30V Max measured current 10A 180 adva...

Страница 17: ...ment result viewing button control and other operations can be achieved through UNI T Smart Measurement APP 3 2 UNI T Smart Measurement APP has multiple functions including Bluetooth communication dat...

Отзывы: