Programming Manual UDP3000S Series
Instruments.uni-trend.com
15
/
49
For Example
:
:SOURce1:VOLTage:PROTection:STATe 1
[:SOURce#]:VOLTage:PROTection:STATe?
Format
:
[:SOURce#]:VOLTage:PROTection:STATe?
Function
:
Query the overvoltage protective switch of a certain channel.
Description
:
The optional value for # are“1|2|3|5|6”. 1 is corresponding to CH1; 2 is corresponding to CH2; 3 is
corresponding to CH3; 5 is corresponding to SER; 6 is corresponding to PARA. If [:SOURce#] or # is
ignored, then the default is CH1. It returns“ON | OFF”.
For Example
:
:SOURce1:VOLTage:PROTection:STATe?
Return
:
ON
[:SOURce#]:CURRent [:LEVel][:IMMediate][:AMPLitude]
Format
:
[:SOURce#]:CURRent [:LEVel][:IMMediate][:AMPLitude] {<curr>|MINimum|MAXimum}
Function
:
Set the current value of a certain channel and set the channel as the current channel.
Description
:
The optional value for # are“1|2|3|5|6”. 1 is corresponding to CH1; 2 is corresponding to CH2; 3 is
corresponding to CH3; 5 is corresponding to SER; 6 is corresponding to PARA. If [:SOURce#] or # is
ignored, then the default is CH1. “1, 2”can only be selected in normal mode; “5” can only be selected in serial
mode; “6”can only be selected in parallel mode.
For Example
:
:SOURce1:CURRent 5.000
[:SOURce#]:CURRent [:LEVel][:IMMediate][:AMPLitude]?
Format
:
[:SOURce#]:CURRent [:LEVel][:IMMediate][:AMPLitude]?
Function
:
Query the current value of a certain channel.
Description
:
The optional value for # are“1|2|3|5|6”. 1 is corresponding to CH1; 2 is corresponding to CH2; 3 is
corresponding to CH3; 5 is corresponding to SER; 6 is corresponding to PARA. If [:SOURce#] or # is
ignored, then the default is CH1.
For Example
:
:SOURce1:CURRent?
Return
:
5.000
[:SOURce#]:CURRent:PROTection[:LEVel]
Format
:
[:SOURce#]:CURRent:PROTection[:LEVel] {<curr>|MINimum|MAXimum}
Function
:
Set the overcurrent protective value of a certain channel and set the channel as the current channel.
Description
:
The optional value for # are“1|2|3|5|6”. 1 is corresponding to CH1; 2 is corresponding to CH2; 3 is
corresponding to CH3; 5 is corresponding to SER; 6 is corresponding to PARA. If [:SOURce#] or # is
ignored, then the default is CH1. “1, 2”can only be selected in normal mode; “5” can only be selected in serial
mode; “6”can only be selected in parallel mode.
For Example
:
:SOURce1:CURRent:PROTection 5.000
[:SOURce#]:CURRent:PROTection[:LEVel]?
Format
:
[:SOURce#]:CURRent:PROTection[:LEVel]?