4.6. Fault
Codes
If a fault is detected during use, the HEAT light will switch off. The LED light L will flash a number of times then
the LED light H will flash a number of times to indicate a code. For example, if the light L flashes 2 times and
then the light H flashes once the fault code is 21.
6
WARNING
The heating element is switched off automatically except in the case of
fault code 22.
Press the OFF button to stop the fault code, follow the required operation as described below and press ON
before starting after the fault has been rectified.
Fault code Flashing LED light
Description of the fault Required Action
Ready L
Ready H
11
1x
1x
The thermocouple of
the tank is broken and
has an open circuit.
Contact the service/
repair department of Uni-ram
Corporation
12
1x
2x
The thermocouple of
the condenser has an
open circuit.
Contact the service/
repair department of Uni-ram
Corporation
13
1x
3x
Computer fault.
The computer must be replaced.
Contact the service/
repair department of Uni-ram
Corporation
21
2x
1x
The heating circuit has
an open circuit.
Contact the service/
repair department of Uni-ram
Corporation
22
2x
2x
The heating triac is
broken and the heating
is still "ON".
Unplug the power cord safely.
Allow the URS600CE to cool down
for at least 3 hours. Under no
circumstances should you open the
lid until the URS600CE has cooled
down completely.
Contact the service/
repair department of Uni-ram
Corporation
23
2x
3x
The condenser is
overheated.
The heating circuit is turned off
automatically until the temperature
in the condenser falls below 75°C.
If this process takes longer than 10
minutes, the URS600CE switches to
a fault code 24.