USER MANUAL - URS600CE AND URS600CESS JULY 2019
19
Clamp handle
Tank lid
Figure 7. Opening the tank lid
2. Check that the tank is empty.
3. Check that the liner bag is suitable for use with no risk of leakages.
4. Install the liner bag so that the bottom is flat on the bottom of the tank.
5. Insert your fingers into the loop of the frame ring and press to reduce the diameter.
6. Insert the frame ring into the liner bag in the tank and release so that the ring is just above the edge in the distillation tank.
7. Fold the lip of the bag all the way around over the ring.
Please note!
Ensure that the liner bag does not block the vapour outlet fitting of the tank.
You have now placed the liner bag.
step 5
step 6
step 7
step 8
Figure 8. Steps for placing the liner bag with the bag holder
3.4. Filling the URS600CE with contaminated solvent
1. Check whether the contaminated solvent meets the requirements as stated in section 3.2.
2. Carefully pour the contaminated solvent into the liner bag in the distillation tank.
Please note!
The URS600CE has a maximum capacity of 20 litres, do not overfill. Fill the tank with at
least 6 litres of the solution.
3. Fill the tank to a maximum of 4-5 cm below the retainer ring, otherwise the contaminated solvent may