flow into the condenser and cause a blockage or stain and / or contaminate the distilled solvent.
Please note!
When pouring in the contaminated solvent, ensure that everything goes into the liner bag
so that no contaminated solvent flows between the liner bag and the distillation tank.
6
BE CAREFUL
Avoid inhalation of vapour during the filling or cleaning of the distillation ttank.
4. Close the tank lid.
5. Lock the tank lid with the lid clamp.
Please note!
Make sure that the tank lid closes properly to prevent leakages.
6. Close the safety cover.
7. Open the door on the front of the URS600CE.
8. Place the supplied HDPE receiving pail with a 3.0 cm diameter opening inside the cabinet of the URS600CE.
Please note!
Always use the supplied HDPE pail with the
3.0 cm
opening. The small opening limits
the possible formation of vapour to a minimum.
Please note!
Do not recycle paint remover and cleaning solvents in the same receiving container.
Please note!
The HDPE pail comes with a cap. Keep the cap in a safe location and return the cap to
the opening of the pail when transporting the solvent in the pail away from the URS600CE.
Figure 9. Place the HDPE receiving pail in the cabinet. Remove the
cap. Position the discharge tube in the 3.0 cm opening.
9. Place the empty HDPE receiving pail with the 3.0 cm opening so that the 13 mm (diameter) condenser outlet hose sticks into
the opening of the pail lid.
Please note!
Make sure the condenser outlet hose is not kinked or blocked and that it extends at least 2.5 cm into the
opening.
10. Close the door on the front of the URS600CE.
You are now ready to set up the URS600CE.