B
Warning Label
Warning
Check the
capacity
Primary power supply
Please prepare the following: Single-phase 100 V, 6.3 A
Smaller power supply capacity may cause the breaker to trip due to
overcurrent during the operation.
Do not share
For the primary power supply, a single power supply
should be provided for the device, and other equipment
should not be connected with it.
Breaker capacity shortage may cause the breaker to trip due to
overcurrent during the operation.
Ground the
grounding wire
Employ Level D grounding.
Use a plug with grounding wire for 100-V power supply.
If you use a plug adapter, connect the grounding wire
with a nearest grounding terminal.
If incorrect grounding is made, this may cause electrical shock in
case of failure or current leakage.
Check the
cable capacity
Avoid using an extension cable. However, if it must be
used by necessity, use the following cable.
For 100-V power supply: 1.25 mm
2
or more
If a thinner cable is used, this may cause overheating, ignition, or
fire.
Avoid this
action
Do not place any object on the cable for the primary side.
Otherwise, such action may cause electrical shock or fire.
Avoid electrical
shock
Do not touch the terminal block or other connectors if the
primary cable plug is being inserted in a socket.
Otherwise, the operator may suffer electrical shock.
Caution
Check labels
The warning labels are pasted on the following parts.
1) Left back of the rack (PL004)
2) Main valve handle (PL008)
3) Around the oil diffusion pump (D. P) (PL007)
When accompanied by a liquid nitrogen trap.
4) Liquid nitrogen trap (PL006)
Contact us if the label is contaminated or peeling off.
Power Supply