4.8 An inlet port, exhaust port replacement (an option)
4.8.1 The switching method of the inlet port
When I change an inlet port of the pump, please choose 40
□
flange of our optional inlet port.
The switching method excludes screw fixing an inlet port, and please attach an inlet port to
change.
4.8.2 The switching method of the exhaust port
When I change exhaust port of the pump, please choose G1 size screw of our optional exhaust
port. The switching method removes exhaust port fixing, and please attach an exhaust port to
change.
4.9 Installation of the oil mist trap (Option)
An oil mist trap can be installed in order to remove oil mist from the pump. As such an oil mist
trap, models OMT-100A and OMI-100 are available. Remove the standard exhaust port from the
exhaust port of the pump and install an oil mist trap instead. The oil mist trap not only prevents
oil mist generation but also reduces exhaust noise.
For details, refer to the instruction manual for the oil mist trap.
4.10 Restriction on operation when the oil mist trap is installed
When using the oil mist trap, there are some restrictions on operation. When the filter is clogged,
replace it.
The internal pump pressure is limited to 0.03 MPa (gauge pressure). When the pressure
measured at the outlet side has reached 0.03 MPa (gauge pressure), replace the oil mist trap
filter.
Be sure to observe the restrictions on operation when the oil mist trap is
installed. There is a risk of explosion. When the filter is clogged, replace
it.