U-Series Operating Manual
12
C6000610
Technical Information
The need to degas
In order to allow optimum ultrasonic activity, the gases present in ordinary
tap water need to be driven out of the cleaning solution.
The time needed to degas the liquid varies depending on the amount of gas
present in the liquid and the quantity of water in the tank. Ultrawave
recommends a degas period of at least 10 minutes.
During the degas cycle, you will see bubbles of gas forming on the inside of
the bath, and slowly rising to the surface. Degassing is complete when you
can no longer see these bubbles.
Another indication of increased “cold
-
boiling” at the
liquid surface indicates that the liquid is degassed.
Once the liquid is degassed, the bath is ready for use.
The effect of heat
Heating the liquid in the bath
will aid the cleaning process.
Normally a temperature of
between 30 to 60°C is sufficient
to accelerate the process. You
will see from the graph below
that optimum cleaning will be
obtained at 60°C
If you are using your bath to
clean medical equipment, it is recommended to limit the fluid temperature
to 40°C. This will avoid the coagulation of protein and other bio burden which
can then bake onto the equipment.
Ultrasonic activity itself will heat up the liquid at a rate of approximately 10-
15°C per hour if in continuous use.
Cleaning time
Cleaning time will depend on application, type and amount of contamination.
General light contamination should be removed in less than 10 minutes.
An indication of cleanliness is when stains are no longer visible, and
contamination no longer appears in a stream from the item being cleaned.
Average power
0
20
40
60
15
25
35
45
55
65
75
85
Temp (Deg C)
P
o
w
e
r
(W
a
tt
s
/I
n
c
h
^
2
)
Av Power
-3dB point
Содержание U Series
Страница 1: ...C6000910 01 issued October 2021 U Series Operator Instruction Manual...
Страница 19: ......