97
NT UF 811 CO GB (c) / March 2014
IV.2 Reading of N words (with N ≤ 125)
The function code for reading N words is 3.
Request
Data
Slave no.
3
address 1st word
number of words
CRC16
Size
1 byte
1 byte
2 bytes
2 bytes
2 bytes
Response
Data Slave no.
3
number of bytes read value 1st word ….. value last word CRC16
Size 1 byte 1 byte
1 byte
2 bytes
…..
2 bytes
2 bytes