www.ultraoptics.com
page 24
Replacing the Spindle Motor Assembly
1. Remove the suction cup.
2. Disconnect the spindle motor power cable from the back of the spindle.
3. Disconnect the vacuum tubing from back of the spindle.
4. Loosen the bolt on the front tip of the spindle swing arm.
5. Pull old assembly out through the top of the arm.
6. Remove the vacuum elbow.
7. Reattach the vacuum elbow onto the new assembly.
a. Add Teflon tape to threads if needed.
8. Slide the new assembly in to the arm making sure to line up the bottom of the vacuum
elbow with the bottom of the spindle swing arm.
9. Tighten the bolt on the front tip of the spindle swing arm.
10. Reconnect the vacuum tubing to the back of the spindle.
11. Reconnect the spindle motor power cable on the back of the spindle.
12. Reattach the suction cup.
Air Regulator Adjustment
1. Locate regulator on back panel
2. If an adjustment needs to be made, pull down
on the regulator adjustment cap and turn until
the air gauge reads 80 PSI.
3. Push adjustment cap back up to lock in place.
Figure 12 - Location of air regulator