www.ultraoptics.com
page 31
31. After selecting Execute, Select Yes to confirm
32. Repeat steps 9-22 for Point 03 and Point 02 (if machine is left to right) or Point 25 and
Point 24 (if machine is right to left).
33. Select AboveLens1A from the Destination drop menu
34. Remove the tray and calibration tools from Tray Position 1 and place in Tray Position 2
and repeat calibration steps
Calibrating the Cognex barcode scanner
From time to time it may be necessary to re-calibrate the barcode scanner. This may result
from the scanner getting bumped out of position, vibration issues, etc. If there are issues
with the scanner reading properly, please follow this procedure to be sure it is properly
configured and calibrated:
1. Open the current version of Dataman software (provided on initial machine start-up and
training).