The handpiece should not be cold sterilized or submerged in any liquid as it will damage the handpiece. 

Handpiece maintenace 

The small o-ring on the handpiece should be inspected on a regular basis. If the o-ring is damaged it 
should be replaced. The small o-ring should be replace every six month, or every two to three months if 
the handpiece is regularly autoclaved. An installation tool and three extra o-rings are provided in the 
handpiece box. Additional o-rings may be ordered using Part # 54138. The large o-ring should not require 
replacement during normal service. If the large o-ring should become damaged a replacement may be 
obtained using Part # 54136. 

Warranty 

Satelec piezo scaler are covered under warranty for 1 year following the date of purchase against defects 
in material or worksmanship. A copy of an invoice is required as proof of date of purchase. Warranty is 
void if it is determined that the product has sustained malicious or delberate damage. Tips have a 90 day 
limited warranty. 

                         

 

 

 Piezo Scaler Quick Reference Guide 

 

Fault Detected

 

Possible Causes

 

Solution

 

Handpiece does not 

vibrate 

1. 

Faulty handpiece cord & connector 

2. 

Scaler cord wire cut 

3. 

Handpiece damaged 

4. 

Electrical contacts on cord are loose 

5. 

Tip worn or distorted 

1. 

Unplug the scaler unit and wait 30 seconds to reset the scaler module 

2. 

Replace tip, worn tips will not allow the scaler to function properly 

3. 

Visi

www.UltimaDentalSystems.com

 to watch the instructional video on how 

to remove the scaler you’re your Ultima 500 series dental unit. Contact the 
UDS Technical support team for further instructions on where to send the 
scaler in for repair (888) 900-8584 

Weak Vibration 

       1.  Tip worn or distorted 
       2.  Tip over tighten 
       3.  Incorrect use 
       4.  Fluid in the cord connector 

       1.  Replace the tip/Check tip chart to see if worn 
       2.  Tighten the tip using the tip wrench 
       3.  Dry electrical contacts 

Fluid leakage between 

hand piece and cord 

1. 

Worn small water o-ring 

2. 

Worn large handpiece o-ring 

1. 

Check the small water o-ring on the bottom of the handpiece near the 
connector for damage. Replace if needed using Part # 54138 

2. 

Check the large handpiece o-ring on the bottom of the handpiece on the 
outer perimeter. Replace the o-ring if needed using Part # 54136 

No light 

(L.E.D model only)

 

1. 

Faulty L.E.D ring 

2. 

Faulty handpiece/cord connector contacts 

1. 

Clean the L.E.D ring contacts/ Replace L.E.D ring 

2. 

Clean the hand piece / cord connector contacts 

3. 

For further assistance call the UDS Technical support team: (888) 900-8584 

Standard Satelec Handpiece 

– Part # 54086 

LED Satelec Handpiece 

– Part # 54082 

Содержание 250 Series

Страница 1: ...www UltimaDentalSystems com 888 900 8584 ULTIMA 250 SERIES Owners Manual...

Страница 2: ...humb screw located at the top of the unit The two suction feet provided are to ensure more stability for the unit on a table top surface even if the water bottles are empty To make a more permanent in...

Страница 3: ...a wall beam Use a stud finder which can be found in most hardware stores to locate a beam in the area the Ultima 250 will be installed However if this is not possible a second option is to use Drywall...

Страница 4: ...re attach the other half tighten the screws on the pole mount until tight and secure Both Table Top and Arm units include a standard 12 arm attachment Additional 6 8 12 and 16 arm attachments are als...

Страница 5: ...female quick disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the compressor to 70 Psi NEVER EXCEED 80 Psi GOING INTO THE UNIT If you are using a compresso...

Страница 6: ...the male quick disconnect fitting with the shut off valve to the female disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the nitrogen source to 70 Psi NEVER...

Страница 7: ...ISTILLED WATER AFTER USE FOR THIRTY SECONDS TO A MINUTE TO NOT ALLOW THE CHOLHEXIDINE TO AND FORM PARTICLES INSIDE THE TUBING LINES WHICH CAN CAUSE CLOGS AND BLOCKAGE WITHIN THE UNIT 4 The white cap n...

Страница 8: ...he holes are small adjustment screws To decrease individual air line pressure turn adjustment screw clockwise To increase individual air line pressure turn counter clockwise To install hand pieces sim...

Страница 9: ...e very little maintenance 1 Always keep handpieces oiled with liquid handpiece oil 2 Disconnect air source going into the unit each night to allow the unit to vent 3 Always remember to flush out the t...

Страница 10: ...transducer The transducer consists of ceramic disks stacked next to one another These ceramic disks have high quartz content and when excited with electricity they compress and decompress causing a co...

Страница 11: ...oesn t heat up so only the tip requires cooling reducing the amount of water used tremendously Adjusted the Water Flow 1 Screw on the 1 tip Universal onto the handpiece until it is hand tight Then use...

Страница 12: ...off water when using a tip without a water port Handpiece and tubing If using irrigating fluids other than water the cord and handpiece should be flushed out with distilled water for at least one minu...

Страница 13: ...Handpiece damaged 4 Electrical contacts on cord are loose 5 Tip worn or distorted 1 Unplug the scaler unit and wait 30 seconds to reset the scaler module 2 Replace tip worn tips will not allow the sca...

Страница 14: ...ized the bottles should be rock hard If no water continues to be the issue the problem may be more involved and may require a professional dental equipment technician Other causes of no water issues c...

Страница 15: ...properly onto the polisher nose cone 1 Faulty or damaged nose cone chuck 2 Prophy angle stem has broken inside of the nose 1 The nose cone chuck will need to be repaired or replaced Contact Ultima Den...

Страница 16: ...side down for thirty seconds to allow the oil to travel through the polisher It is important that the polisher be run upside to avoid the thin handpiece oil from sliding back down the handpiece tubing...

Страница 17: ...drill is most likely damaged and needs to be replaced 1 Remove the end cap located on the back of the drill Check for debris this is a sign that parts inside the turbine such as the bearings may have...

Страница 18: ...of the two larger barbs located on the bottom of the drill Re attach the hand piece and run the drill upside down for thirty second to a minute to allow the oil to circulate through the drill Turning...

Страница 19: ...ent spare parts for T64 highspeed drills are still available for purchase If you have any questions or need assistance ordering parts for T64 highspeed drills contact Ultima Dental Systems at 888 900...

Страница 20: ...warranty if defective out of the box and damaged during shipping All parts on Ultima compressors are covered for 5 years if defective The Lowspeed Polisher is covered under warranty for 1 year if defe...

Отзывы: