Basic Trouble Shooting Guide 

 

No water - Scaler/High-speed Drill 

 

1.  Check the water on/off toggle switch and make sure it is in the on position. 
2.  Check the water increase/decrease knob and make sure it is turned up. 
3.  Grab the 3-way syringe and check to see if you get water from the handpiece. 
4.  Inspect the water bottles. Check for air leaks around the water bottles and make sure their 

secured tightly to the pressure head.  

5.  Squeeze the bottles to make sure they pressurized; the bottles should be rock hard. 

 

 

If  no  water  continues  to  be  the  issue,  the  problem  may  be  more  involved  and  may  require  a 
professional  dental  equipment  technician.  Other  causes  of  no  water  issues  could  involve  the 
blockage,  failure,  or  damage  of  the  water  relay,  control  block,  control  block  diaphragm,  or  the 
distribution block.   

To  determine  the  exact  cause  of  the  issue  and  to  schedule  an  in-office  visit  by  a  certified 
technician Call Pro Repair Service: 1-800-645-6594. For all other general questions contact the 
UDS Technical Support team: (888) 900-8584.  

Please check with the local dental Repair Service or equipment distributor in your area if you wish 
to schedule an in-office visit outside of the U.S.A. 

 

 

 

 

 

 

 

Содержание 250 Series

Страница 1: ...www UltimaDentalSystems com 888 900 8584 ULTIMA 250 SERIES Owners Manual...

Страница 2: ...humb screw located at the top of the unit The two suction feet provided are to ensure more stability for the unit on a table top surface even if the water bottles are empty To make a more permanent in...

Страница 3: ...a wall beam Use a stud finder which can be found in most hardware stores to locate a beam in the area the Ultima 250 will be installed However if this is not possible a second option is to use Drywall...

Страница 4: ...re attach the other half tighten the screws on the pole mount until tight and secure Both Table Top and Arm units include a standard 12 arm attachment Additional 6 8 12 and 16 arm attachments are als...

Страница 5: ...female quick disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the compressor to 70 Psi NEVER EXCEED 80 Psi GOING INTO THE UNIT If you are using a compresso...

Страница 6: ...the male quick disconnect fitting with the shut off valve to the female disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the nitrogen source to 70 Psi NEVER...

Страница 7: ...ISTILLED WATER AFTER USE FOR THIRTY SECONDS TO A MINUTE TO NOT ALLOW THE CHOLHEXIDINE TO AND FORM PARTICLES INSIDE THE TUBING LINES WHICH CAN CAUSE CLOGS AND BLOCKAGE WITHIN THE UNIT 4 The white cap n...

Страница 8: ...he holes are small adjustment screws To decrease individual air line pressure turn adjustment screw clockwise To increase individual air line pressure turn counter clockwise To install hand pieces sim...

Страница 9: ...e very little maintenance 1 Always keep handpieces oiled with liquid handpiece oil 2 Disconnect air source going into the unit each night to allow the unit to vent 3 Always remember to flush out the t...

Страница 10: ...transducer The transducer consists of ceramic disks stacked next to one another These ceramic disks have high quartz content and when excited with electricity they compress and decompress causing a co...

Страница 11: ...oesn t heat up so only the tip requires cooling reducing the amount of water used tremendously Adjusted the Water Flow 1 Screw on the 1 tip Universal onto the handpiece until it is hand tight Then use...

Страница 12: ...off water when using a tip without a water port Handpiece and tubing If using irrigating fluids other than water the cord and handpiece should be flushed out with distilled water for at least one minu...

Страница 13: ...Handpiece damaged 4 Electrical contacts on cord are loose 5 Tip worn or distorted 1 Unplug the scaler unit and wait 30 seconds to reset the scaler module 2 Replace tip worn tips will not allow the sca...

Страница 14: ...ized the bottles should be rock hard If no water continues to be the issue the problem may be more involved and may require a professional dental equipment technician Other causes of no water issues c...

Страница 15: ...properly onto the polisher nose cone 1 Faulty or damaged nose cone chuck 2 Prophy angle stem has broken inside of the nose 1 The nose cone chuck will need to be repaired or replaced Contact Ultima Den...

Страница 16: ...side down for thirty seconds to allow the oil to travel through the polisher It is important that the polisher be run upside to avoid the thin handpiece oil from sliding back down the handpiece tubing...

Страница 17: ...drill is most likely damaged and needs to be replaced 1 Remove the end cap located on the back of the drill Check for debris this is a sign that parts inside the turbine such as the bearings may have...

Страница 18: ...of the two larger barbs located on the bottom of the drill Re attach the hand piece and run the drill upside down for thirty second to a minute to allow the oil to circulate through the drill Turning...

Страница 19: ...ent spare parts for T64 highspeed drills are still available for purchase If you have any questions or need assistance ordering parts for T64 highspeed drills contact Ultima Dental Systems at 888 900...

Страница 20: ...warranty if defective out of the box and damaged during shipping All parts on Ultima compressors are covered for 5 years if defective The Lowspeed Polisher is covered under warranty for 1 year if defe...