Basic Trouble Shooting Guide 

Polisher 

 

Fault Detected 

Possible Causes 

Solution 

The prophy angle  

does not spin 

1. 

The prophy is faulty or damged 

2. 

The gear dial is set to the incorrect 
position 

3. 

The polisher motor is ceased 

1. 

Replace the prophy angle. The bearings on metal prophy 
angles often go bad and need to be replaced. We recommend 
using Crosstex disposable prophy angles to prevent costly 
repairs (Part # UD0311). 

2. 

Make sure the polisher motor is set to the correct gear. If the 
motor is set to the neutral position, the prophy angle will not 
spin. (See Picture Below) 

3. 

Try oiling the polisher to break free the ceased motor. Watch 
the instruction video a

www.UltimaDentalSystems.com

 

The prophy angle 

 will not seat properly onto 

the polisher nose cone 

1. 

Faulty or damaged nose cone chuck 

2. 

Prophy angle stem has broken inside of 
the nose 

1. 

The nose cone chuck will need to be repaired or replaced. 
Contact Ultima Dental Systems to send in your handpiece in 
for repair and receive a loaner handpiece while you wait, call 
(888) 900-8584. 

2. 

The prophy angle stem needs to be removed from the nose 
cone chuck, this requires disassembling the polisher nose 
cone. Contact Ultima Dental Systems to send in your 
handpiece in for repair and receive a loaner handpiece while 
you wait, call (888) 900-8584.  

The polisher seems weak 

and lacks power 

1. 

The pressure is not set correctly for the 
polisher handpiece 

1. 

Polisher  pressure should be  between  40-50 Psi.   Adjust  gold 
screw  that  reads  HP  1  on  the  front  panel  if  needed.  When 
adjusting, the pressure turn the screw slowly to avoid damaging 
the control block diaphragm. 
 

 

 

 

 

 

 

 

 

         

     

 

   

 

 

 

 

 

 

 

 

Gear Lines 

Gear Indicator

Prophy Angle Indicator 

Collar Up 

Collar Down 

Содержание 250 Series

Страница 1: ...www UltimaDentalSystems com 888 900 8584 ULTIMA 250 SERIES Owners Manual...

Страница 2: ...humb screw located at the top of the unit The two suction feet provided are to ensure more stability for the unit on a table top surface even if the water bottles are empty To make a more permanent in...

Страница 3: ...a wall beam Use a stud finder which can be found in most hardware stores to locate a beam in the area the Ultima 250 will be installed However if this is not possible a second option is to use Drywall...

Страница 4: ...re attach the other half tighten the screws on the pole mount until tight and secure Both Table Top and Arm units include a standard 12 arm attachment Additional 6 8 12 and 16 arm attachments are als...

Страница 5: ...female quick disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the compressor to 70 Psi NEVER EXCEED 80 Psi GOING INTO THE UNIT If you are using a compresso...

Страница 6: ...the male quick disconnect fitting with the shut off valve to the female disconnect fitting on the side of the unit labeled Filtered Air In Set the air pressure from the nitrogen source to 70 Psi NEVER...

Страница 7: ...ISTILLED WATER AFTER USE FOR THIRTY SECONDS TO A MINUTE TO NOT ALLOW THE CHOLHEXIDINE TO AND FORM PARTICLES INSIDE THE TUBING LINES WHICH CAN CAUSE CLOGS AND BLOCKAGE WITHIN THE UNIT 4 The white cap n...

Страница 8: ...he holes are small adjustment screws To decrease individual air line pressure turn adjustment screw clockwise To increase individual air line pressure turn counter clockwise To install hand pieces sim...

Страница 9: ...e very little maintenance 1 Always keep handpieces oiled with liquid handpiece oil 2 Disconnect air source going into the unit each night to allow the unit to vent 3 Always remember to flush out the t...

Страница 10: ...transducer The transducer consists of ceramic disks stacked next to one another These ceramic disks have high quartz content and when excited with electricity they compress and decompress causing a co...

Страница 11: ...oesn t heat up so only the tip requires cooling reducing the amount of water used tremendously Adjusted the Water Flow 1 Screw on the 1 tip Universal onto the handpiece until it is hand tight Then use...

Страница 12: ...off water when using a tip without a water port Handpiece and tubing If using irrigating fluids other than water the cord and handpiece should be flushed out with distilled water for at least one minu...

Страница 13: ...Handpiece damaged 4 Electrical contacts on cord are loose 5 Tip worn or distorted 1 Unplug the scaler unit and wait 30 seconds to reset the scaler module 2 Replace tip worn tips will not allow the sca...

Страница 14: ...ized the bottles should be rock hard If no water continues to be the issue the problem may be more involved and may require a professional dental equipment technician Other causes of no water issues c...

Страница 15: ...properly onto the polisher nose cone 1 Faulty or damaged nose cone chuck 2 Prophy angle stem has broken inside of the nose 1 The nose cone chuck will need to be repaired or replaced Contact Ultima Den...

Страница 16: ...side down for thirty seconds to allow the oil to travel through the polisher It is important that the polisher be run upside to avoid the thin handpiece oil from sliding back down the handpiece tubing...

Страница 17: ...drill is most likely damaged and needs to be replaced 1 Remove the end cap located on the back of the drill Check for debris this is a sign that parts inside the turbine such as the bearings may have...

Страница 18: ...of the two larger barbs located on the bottom of the drill Re attach the hand piece and run the drill upside down for thirty second to a minute to allow the oil to circulate through the drill Turning...

Страница 19: ...ent spare parts for T64 highspeed drills are still available for purchase If you have any questions or need assistance ordering parts for T64 highspeed drills contact Ultima Dental Systems at 888 900...

Страница 20: ...warranty if defective out of the box and damaged during shipping All parts on Ultima compressors are covered for 5 years if defective The Lowspeed Polisher is covered under warranty for 1 year if defe...

Отзывы: