![UINICS SP-564A Series Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/uinics/sp-564a-series/sp-564a-series_instruction-manual_3284762032.webp)
- 31 -
16. Adjusting the analog output
The analog output(AV3 - 5,AI)range is adjusted correctly at a factory.
Please do not touch except necessity.
・Tune in exactly
1. Power on Mode key being pressed to put the instrument into the test mode.
2. Press the Mode key until the analog output test “dA” appears.(Refer to page 12.)
3. It adjusts so that it may become the output value corresponding to indication value
as follows. Be sure to start the adjustment from the zero volume.
For the voltage output(AV3)type unit
Indication
Output voltage
0
1 V
Turn the zero volume to adjust
10
5 V
Turn the span volume to adjust
(※ Repeat the procedure several times for fine adjustment.)
For the voltage output(AV4)type unit
Indication
Output voltage
0
0 V
Turn the zero volume to adjust
10
5 V
Turn the span volume to adjust
(※ Repeat the procedure several times for fine adjustment.)
For the voltage output(AV5)type unit
Indication
Output voltage
0
0 V
Turn the zero volume to adjust
10
10 V
Turn the span volume to adjust
(※ Repeat the procedure several times for fine adjustment.)
For the current output(AI)type unit
Indication
Output current
0
4 mA
Turn the zero volume to adjust
10
20 mA
Turn the span volume to adjust
(※ Repeat the procedure several times for fine adjustment.)
(Fig 21)