23

valid measuring value. Insert the proper value and press OK.

When enabled, the measurement menu will show as follows:

Measurement Menu

The

 “AUTO” icon

 will show as orange, meaning that the device is ready to 

measure. Press the 

“ZERO” button

 on the screen.

To enable the measuring phase, press the PLAY button or press the switch pedal. 

The 

“AUTO” icon

 will turn green.

Dip the animal paw into the cell, paying attention to the position of the paw in the 

cell.
When the MINIMUM VOLUME and MAXIMUM FLUCTUATION are met, after the 

STABILIZATION TIME the measure is automatically stored in the memory.

6.4.  Restoring Liquid Level

The water tank has been provided for easing the task of liquid level restoring, 

which may be necessary after having processed 20-30 animals.
Consider that each paw takes away some drops of liquid when withdrawn from 

the cell.
It is also recommended to empty and clean and refill the cell in case of droppings 

from the animal.

Back to Contents

Содержание 37240

Страница 1: ...YOUR COMPANION IN DISCOVERY SINCE 1963 Plethysmometer User Guide SKU 37240 Rev 1 0 Nov 2023 ugobasile com...

Страница 2: ...tions Service and adjustments should be carried out by qualified personnel authorized by Ugo Basile organization Any adjustment maintenance and repair of the powered instrument should be avoided as mu...

Страница 3: ......

Страница 4: ...1 Setting Up the Plethysmometer 14 6 1 1 Preparing the dipping solution 14 6 1 2 Main Menu 15 6 1 3 Utilities Menu 16 6 1 4 Device setup menu 18 6 2 Calibration 20 6 2 1 How to calibrate 20 6 3 Measu...

Страница 5: ...mily feeling design Displays the exact paw volume on the touch screen display read out with resolution of 0 01ml Small differences are detected by an high precision self made level transducer Foot ped...

Страница 6: ...101 1 USB Pen Drive including instruction manual with CE certi fication and X Pad software 37215 303 1 Pedal Switch Plethysmometer kit with 25mm Rat paw tube cell SKU 37240 25 SKU Q ty Description 372...

Страница 7: ...n manual with CE certi fication and X Pad software 37215 303 1 Pedal Switch Optional items ordering information SKU Description 37252 Standard water cell with Mouse paw tube diam 13 mm and Rat paw tub...

Страница 8: ...the rat paw 7187 25 mm tube for the rat paw 7189 35 mm tube for the rat paw 7153 L Water level transducer E BT 008 Battery type CR2032 E FT010 1 2 X fuse T1 25A 6X32 mm Consumables SKU Description 716...

Страница 9: ...transducer of original design which measures small differences in liquid level caused by volume displacement is connected to an electronic unit used to show save and recall the values 3 1 Principle o...

Страница 10: ...our digits with 0 01 ml resolution 4 2 The water cell The standard kit SKU 37240 includes a standard water cell for Rat diam 18 mm provided with an extra paw tube diam 13 mm for measuring the mouse pa...

Страница 11: ...or the correct installation and operation of the instrument Please save the manual in a safe place ready to be consulted by qualified personnel who use the instrument Print it only if necessary Our Us...

Страница 12: ...eld responsible for damage caused to things and people and warranty will be void due to Incorrect electrical supply Incorrect installation procedure Incorrect or improper use or in any case not in acc...

Страница 13: ...al for details can be uploaded into the device by using a USB pen see paragraph 4 2 DO NOT PLUG 2 DEVICES AT THE SAME TIME The lower USB port is not available please DO NOT REMOVE THE CAP TTL I O 15pi...

Страница 14: ...d rotation of the transducer head to make sure it is properly fitted to the smaller tube Fill the cell by opening the loading valve by turning its wing horizontally Keep the unloading tap closed while...

Страница 15: ...r with the Plethysmometer home screen is very easy see figure Main Menu Main Menu From the home page press the Ugo Basile logo button on the top bar for software and memory information Press the back...

Страница 16: ...are revision and the serial number of your device asking for an update In order check out your actual device firmware revision press the Ugo Basile logo icon at the home page windows and note the revi...

Страница 17: ...mited we advice you to perform this function after You saved Your experimental data on your PC at the end of a experiment session WARNING all the Experiment and result data will be lost with no way to...

Страница 18: ...IP assignment use auto DHCP Dynamic Host Configuration Protocol to let the device IP address being automatically assigned When connecting the device to a facility LAN make sure with IT department a DH...

Страница 19: ...ected devices Gateway this data is not actually in use since the device does not access the Internet Leave it 0 0 0 0 Primary DNS in DHCP mode this data is automatically populated but it is not needed...

Страница 20: ...ce only 0 2 ml orange for reference only 0 5 ml pink 1 ml grey 2 ml blue 4 ml green Calibration is a linearisation of the relationship between the transducer measure and the actual volume displacement...

Страница 21: ...is calibration volumes plus the volume of these drops Also space the calibration by at least 10 second intervals either when calibrating or evaluating the oedema This will let the liquid film formed o...

Страница 22: ...3 2 Automatic measurement To make the measurement easier and let the operator concentrate on dipping the paw in a most precise way there is a semi automatic measurement function that needs to be setup...

Страница 23: ...al paw into the cell paying attention to the position of the paw in the cell When the MINIMUM VOLUME and MAXIMUM FLUCTUATION are met after the STABILIZATION TIME the measure is automatically stored in...

Страница 24: ...lues volume ml calibration value mV ml and zero value mV in addition to the descriptive data if they have been input in the Experiment field Treatment Protocol ID Stage Trial See image below Result Me...

Страница 25: ...transfer data unit This feature can be useful for example when lab rules does not permit the use of a USB pen drive for security reasons Requirements To use this feature it must be ensured that The d...

Страница 26: ...ill the next field New Password and confirm it filling the Repeat new password field Note that Password length must be between 8 and 15 characters and should contain at least a lower case letter an up...

Страница 27: ...hysmometer device then log in From the menu go to Experiment and press Choose File this text is depending the operative system language point to the X PAD file and press Send to device button Experime...

Страница 28: ...TL OUT Reserved 2 Reserved TTL OUT Reserved 3 Reserved TTL OUT Reserved 4 Reserved TTL OUT Reserved 5 Reserved TTL OUT Reserved 6 Reserved TTL OUT Reserved 7 Reserved TTL OUT Reserved 8 Reserved ANALO...

Страница 29: ...anical stiffness and perfect water tightness without any clamping or fastening device To detach the tubes in case replacement becomes necessary exert pull and rotation as when uncorking a bottle To in...

Страница 30: ...our local distributor either directly or via our support web form Customer Support Contacts Phone 39 0332 744574 E mail service ugobasile com Web form https ugobasile com support support request Befor...

Страница 31: ...0 UBC12 UB Care 12 Additional hardware warranty extension 12 months for Plethysmometer Valid for SKU 37240 37240 UBC24 UB Care 24 Additional hardware warranty extension 24 months for Plethysmometer Va...

Страница 32: ...ts of lymphedema in the hindlimb of mice Scientific Reports Y Fotio O Sasso R Ciccocioppo D Piomelli 2021 Antinociceptive Profile of ARN19702 2 Ethylsulfonylphenyl 2S 4 6 fluoro 1 3 benzothiazol 2 yl...

Страница 33: ...on D Amour Smith method 38450 Electronic Von Frey e VF Handheld 37450 275 Von Frey Filaments manual touch sensitivity kit Estimates suggest that 20 of adults suffer from pain globally Chronic pain is...

Страница 34: ...of thermal hyperalgesia allodynia 38500 PAM Pressure Application Measurement for joint pain 31300 Orofacial Stimulation Test Fehrenbacher Henry Hargreaves method 37300 I R Heat Flux Radiometer for Ta...

Страница 35: ...tations in the main bibliographic search engines Ugo Basile SRL Societ Unipersonale Via Giuseppe Di Vittorio 2 21036 Gemonio VA ITALY Tel 39 0332 744574 sales ugobasile com service ugobasile com Your...

Отзывы: