Operation
22
6.3.1. Manual measurement
The manual measurement mode is shown in the “
MAN
” icon on the right of the
selected paw, highlighted in grey.
Press the
“ZERO” button
on the screen and then dip the animal paw into the cell,
paying attention to the position of the paw in the cell.
Dipping the paw at different levels for repeated measurement may result in
different readings.
We advise You to mark with a soft marker the animal paw to have a reference in
order dipping the animal paw at the same level during repeated readings
There are two modes to save the measured value:
• Press the pedal switch and the value is automatically stored in the internal
memory of the device.
• Press the
“Play” button
on the screen.
6.3.2. Automatic measurement
To make the measurement easier and let the operator concentrate on dipping the
paw in a most precise way, there is a semi-automatic measurement function that
needs to be setup.
From the Main Menu select the “
Device Setup
” then press the
“Auto Reading”
button
. The following menu is shown:
Automatic Measurement Menu
Enable the automatic measurement by pressing the
“Auto Reading” button
. The
button turn to “YES”. The following parameters are described in details:
• MAXIMUM FLUCTUATION: it is the measuring delta within which a certain value
is considered valid during the acquisition phase. Insert the desired value then
press OK.
• STABILIZATION TIME: it is the time that must pass before saving the measured
value when maximum fluctuation and minimum volume parameters are met.
Insert the proper value then press OK.
• MINIMUM VOLUME: it’s the minimum volume displacement that triggers a