background image

Instructions for use (rev 251121) 

e.ON sinter base 

 

UGIN DENTAIRE 

 

10

 

5.

 

START-UP 

After checking that the mains voltage matches that noted on the rating plate, plug the cord into a standard 

20A socket

 with a ground terminal. 

NOTE: If not used for a long time, switch off the power supply. 
 

a)

 

Activate the breaker (ON/OFF) of furnace. 

 

 

 

b)

 

Following power up, the initialization interface appears 
on the screen. The main screen automatically appears 
as follows: 

menu “Working interface”.

 

c)

 

Open the door of the heating chamber.  
Remove  the  protections  and  dust  with  a  vacuum  cleaner  or  a  soft  and  clean  brush  (do  not  use 
compressed air). 
 

 

 

DO NOT PLACE ANY OBJECT INSIDE THE CHAMBER 

(only tray for sinterizing and cover, sole and insulating plate)

 

 

 

 

IMPORTANT 

 

For the first time or after a long time without use, need to

 

preheat the muffle furnace

 1 hour 

at 120°C follow 2 hours at 300°C

 

to eliminate condensation and avoid fractures of the refractory 

material.

 

 

Carry out a decontamination program as a precautionary measure before the first use of the 
furnace

: start the pre-recorded program number 40 

(see chapter 5.6.1).

 

 

 
 

 

Содержание e.on sinter base

Страница 1: ...INSTRUCTIONS FOR USE Sintering furnace R v 251121...

Страница 2: ...NTERFACE 11 6 2 CONTROL KEYS 11 6 3 REAL TIME CURVE CURRENT CYCLE 11 6 4 INTERFACE PROGRAM EDITOR 12 6 4 1 CHANGE SETTINGS 12 6 4 2 PROGRAMMING INSTRUCTIONS 13 6 4 2 1 EXAMPLES OF CALCULATION 14 6 5 P...

Страница 3: ...ter base UGIN DENTAIRE 2 A EQUIPEMENT DESCRIPTION N DESCRIPTION 1 Door of the chamber 2 Magnetic tablet 3 Touch screen 4 Power switch ON OFF lateral side 5 Sole 2 units 6 Insulating plate 7 Tray for s...

Страница 4: ...Instructions for use rev 251121 e ON sinter base UGIN DENTAIRE 3 B WIRING DIAGRAM...

Страница 5: ...e insulating plate tray for sinterizing and cover may reach high temperatures do not touch Wear suitable gloves for handling prosthetic objects tray for sinterizing cover sole and insulating plate The...

Страница 6: ...equipment risk The chamber is made up refractory ceramic fibres Handle carefully They can release dust remove with a vacuum cleaner do not blow or use compressed air Use only genuine spare parts The...

Страница 7: ...must be disposed in accordance with applicable regulations in force 2 EC DECLARATION OF CONFORMITY The equipment s EC declaration of conformity is provided as appendix of instructions for use Content...

Страница 8: ...3 2 TECHNICAL DATA Maximum temperature 1 600 C Temperature accuracy 1 C Heating rate To 40 C minute maximum Screen High definition colour touch screen 7 155x87mm Number of programs 40 free programs w...

Страница 9: ...ments polystyrene Take hold the furnace by its base never lift by upper part of furnace For transport the furnace must remain in an upright position not horizontal position Install the furnace in the...

Страница 10: ...ng term can discolour surfaces that are nearest Do not shake the furnace avoid vibrations and shock The molybdenum silicon rods are very brittle at room temperature DO NOT HANDLE THEM Avoid moving the...

Страница 11: ...menu Working interface c Open the door of the heating chamber Remove the protections and dust with a vacuum cleaner or a soft and clean brush do not use compressed air DO NOT PLACE ANY OBJECT INSIDE T...

Страница 12: ...ppears as shown below Working interface Parameters current cycle display Real time curve Display of the cycle curve temperature time Program editor Program settings Parameter setting Software settings...

Страница 13: ...gram is two pages the first page displays the pairs from C01 T01 to C20 T20 and the second page from C21 T21 to C40 T40 The NEXT PAGE key switches from the first page to the second page 6 4 1 CHANGE S...

Страница 14: ...n of third stage 0 to 9999 minutes Temperature 4 Final temperature of third stage and start temperature of fourth stage If necessary the parameters are programmed as for the previous steps Last time 1...

Страница 15: ...perature remains to 1250 C Time remains 2 hours Time 04 2 hours x 60 min 120 min T04 120 minutes Zirconia SINTERING TEMPERATURES are given as indicative by disc manufacturer If necessary these one can...

Страница 16: ...C10 0 0 C15 0 0 C20 0 0 T05 70 T10 0 T15 0 T20 0 A DECONTAMINATION PROGRAM MUST BE CARRIED OUT AS A PRECAUTIONARY MEASURE BEFORE THE FIRST USE OF THE FURNACE start the pre recorded program number 40 I...

Страница 17: ...l whitish traces still contaminated repeat a decontamination cycle rods with a grey shiny appearance ok furnace ready 6 7 POSITIONING THE ELEMENTS AND STARTING THE CYCLE IMPORTANT Before use check if...

Страница 18: ...the insulating plate vertically for closing the chamber Close the door of the chamber Note the sole can remain at its location for the realization of the following cycles At the end of the cycle wear...

Страница 19: ...displayed on the screen below 300 C When the temperature is below 200 C the power supply can be switched off switch at the back of furnace At the end of the cycle the furnace automatically returns to...

Страница 20: ...pieces absorption of pollutants or a powder specially designed to improve the effectiveness of decontamination After running the program check the condition of the rods still whitish traces still con...

Страница 21: ...creen damage do not use a sharp object to operate the touchscreen In case of dust into the chamber remove it with a vacuum cleaner do not used compressed air Visual inspection of the sole insulating p...

Страница 22: ...heating elements with a wrench supplied as an accessory 3 Repeat step 2 for all the heating elements to replace 4 Unscrew the support elements of heating element with a wrench supplied as an accessory...

Страница 23: ...k the program parameters temperature and time Adjust sintering temperatures if necessary see chapter 6 4 The furnace does not heat up It is necessary to test the continuity of the rods Do not use the...

Страница 24: ......

Отзывы: