1. Failure and error messages
Failure en-
countered
Possible cause
Corrective action
Display Err
- 300
Excessive cuff pressure. Cuff
inflated to maximum due to
movement of the arm or body.
The hose may be kinked or
the air connector blocked.
- Repeat measurement
- Do not move arm
- Do not talk
- Check that hose with air connector is clear
(unobstructed)
Display
Err - 1
Measurement error. Measure-
ment could not be carried
out due to movement of the
arm or body.
- Repeat measurement
- Do not move arm
- Do not talk
Display
Err - 2
Inflation error, pressure built
up too quickly or too slowly.
Check that air connector is correctly fitted,
repeat measurement.
Fault was recognised. Cuff is
not correctly fitted.
- Repeat measurement
- Do not move arm
- Do not talk
Display
Err - 3
Deflation rate too high or
too low.
Keep still during measurement. Call the
Service Hotline if this fault occurs repeatedly.
Display
Err - 4
Microphone error
- Insert connector fully into the unit
- Check for correct fitting of the cuff and the
microphone
- If necessary remove any clothing between
the cuff and the upper arm
- Repeat the measurement
- Rest lower arm flat with palm of hand
facing upwards
If the problem persists, call our Service
Hotline.
Display 0
0
Start/Stop button accidentally
pressed while battery was
being changed.
Switch unit off and then on again with Start/
Stop button; if necessary, activate time signal
reception of radio clock again. See Page
EN-57.
Irregular heartbeat, disrup-
tive movements, shaking,
wobbling (objects), breathing
in deeply, etc.
Repeat measurement after 3-5 minutes rest.
Possible factors (depending on the severity of
the arrhythmia) affecting the measurement
results must be discussed with your doctor.
Failure encoun-
tered
Possible cause
Corrective action
Measured values
/ pulse values are
extremely high
or low
Too little rest before the measure-
ment. Movement or talking during
measurement.
Repeat measurement after
a break of approx. 3-5 mins.
Observe instructions for use
Page EN-55.
Has the proper cuff been select-
ed? Cuffs that are too small result
in excessive blood pressure values.
Measure your arm circumfer-
ence in the middle of the upper
arm and compare this with the
details given on the cuff.
Incorrect measurement position.
Repeat measurement, ensuring
correct body posture. (Page
EN-55)
Consecutive mea-
surements produce
different measured
values (readings).
Blood pressure is not a fixed
quantity. It may vary in the
upward or downward direction
within a few minutes by more
than 20 mmHg.
--------
Measurement was
interrupted and
cuff first deflated
and then inflated
Unit detects a fault or that diasto-
lic value is too low.
Unit shows no fault. Repeat
measurement.
Movement during measurement
If necessary, interrupt measure-
ment, measure again after a
break of 5 minutes.
No display after
unit is switched on
Batteries inserted incorrectly.
Check position of batteries.
Batteries flat.
Change batteries.
Battery contacts dirty.
Clean battery contacts.
Dashes in display
after Memory
button is pressed
No measured values stored.
No display when no measured
values are stored.
Measurement
interrupted
Batteries flat
Change batteries.