Technical information

Technical information

Technical data

No display after switching on. Check that the battery is 
positioned correctly and the polarity (+/-) is correct. If the 
error persists, replace the battery.

Type:

Domotherm

®

 E ear thermometer

Measuring 
range:

0.1 °C to 99.9 °C (32.2 °F to 211.8 °F)

Measurement 
resolution:

0.1 °C/°F

Measurement 
accuracy in the 
laboratory:

± 0.2 °C between 32.0 °C and 42.2 °C 
(± 0.4 °F between 89.6 °F and 108.0 °F) 
± 1 °C (± 2 °F) in other temperature ranges

Display:

Liquid crystal display (LCD), four digits plus special 
icons

Fever alarm:

If temperature values above 37.5 °C are meas-
ured, then three short beeps and a red back-
ground light on the display indicate that this is a 
raised temperature.

Memory:

12 measured values with date and time can be 
opened in memory mode

Operating 
conditions:

Ambient temperature 10 °C to 40 °C (50 °F to 104 
°F), maximum relative humidity 15 % to 95 %

Conditions for 
storage and 
transport:

-25 °C to +55 °C (-13 °F to +131 °F), maximum 
relative humidity 15 % to 95 %

Time to adjust 
from mini-
mum/maxi-
mum storage 
temperature 
to operating 
temperature:

2 hours

Customer service

Metrological inspection

Device repairs may only be performed by the manufacturer or an ex-
pressly authorised body. Please contact:

In accordance with the Medizinprodukte-Betreiberverordnung (Ger-
man Medical Device Operator Ordinance), professional users in Ger-
many are required to arrange for a metrological inspection to be per-
formed at yearly intervals. Private users are advised to arrange for a 
technical inspection of the device every two years.

UEBE Medical GmbH 
Zum Ottersberg 9 
97877 Wertheim, Germany 

Tel.:  +49 (0) 93 42 / 92 40 40 
Fax:  +49 (0) 93 42 / 92 40 80 
Email:  [email protected] 
Website:  www.uebe.com

The inspection can be performed either by UEBE Medical GmbH, an 
authority responsible for metrology or an authorised maintenance 
service. Please refer to your national regulations.

Automatic 
switch-off:

Approx. 30 seconds after last activity

Battery:

1 x CR2032, 3.0 V

Battery life:

Approx. 1000 measurements (with a new battery)

IP classification: IP 22: protected against solid foreign bodies with 

a diameter of 12.5 mm and above, protected 
against dripping water.

Average service 
life:

Five years or 12,000 measurements

Dimensions:

145 mm (L) x 38 mm (W) x 36 mm (H)

Weight:

54 g with battery, 52 g without battery

Содержание Domotherm E

Страница 1: ...Instructions for use www domotherm de...

Страница 2: ...ounding tissue A sensor takes this thermal radiation and converts it into temperature values The measured temperature appears on the display Measurements taken on the surrounding tissue of the ear can...

Страница 3: ...a high level of humidity Read these instructions for use carefully to ensure that you have un derstood all of the functions and the safety information Take the time to familiarise yourself with the t...

Страница 4: ...rowaves or other devices with strong electromagnetic fields can cause mal functions Maintain a distance of at least one metre from such de vices when using the thermometer If you cannot maintain the m...

Страница 5: ...for entering the basic settings date time temperature unit Before using the device for the first time set the date and time so that the measurements can be saved together with the date and time During...

Страница 6: ...ounding tissue of the ear canal will result in lower measured values and can lead to an incorrect diagnosis of a fever For this reason it is very important to position the thermometer in the ear corre...

Страница 7: ...hes back to measuring mode Figure 5 Performing measurement The best way to perform the measurement on small children is to lie them flat on their stomach with their head facing to the side so that the...

Страница 8: ...the normal body temperature varies de pending on where the measurement is taken Caution It is essential to consult a doctor in cases of high fever 37 5 C to 38 0 C high temperature 38 0 C to 39 0 C m...

Страница 9: ...CR2032 lithium battery If the battery icon flashes on the display while the thermometer is in use the battery is low Although you are still able to take tempera ture measurements you should have a sp...

Страница 10: ...Cover of battery compartment unlocked Cover of battery compartment locked Error messages Measured temperature is above 100 0 C and is therefore outside the measuring range Measured temperature is bel...

Страница 11: ...re to operating temperature 2 hours Customer service Metrological inspection Device repairs may only be performed by the manufacturer or an ex pressly authorised body Please contact In accordance with...

Страница 12: ...C 25 C 95 15 IP 22 Disposal Technical equipment and batteries do not belong in household waste They must be disposed of at appropriate collection and disposal points Use one of the enclosed alcohol w...

Страница 13: ...device by the buyer Further more the warranty does not constitute grounds for asserting claims for damages against us UEBE Medical GmbH Service Center Zum Schlag 18 97877 Wertheim Germany Warranty cla...

Страница 14: ......

Страница 15: ...ct to technical modifications Reprinting including of extracts is prohibited Copyright 2019 UEBE Medical GmbH 7 0875 201 DA 2019 07 UEBE Medical GmbH Zum Ottersberg 9 97877 Wertheim Germany Phone 49 0...

Отзывы: