OPERATION
CHARGING THE BATTERY
When the battery indicator
appears in the lower
lefthand corner of the display, the battery needs to be
recharged.
Plug the AC adapter into the scale to charge the battery.
Scale does not need to be turned on. It will take 12 hours
to fully charge the battery.
The charging indicator is located on the bottom left
of the display, to the left of the word RECHARGE. The indicator
light color will change to indicate the battery status:
• ORANGE – Battery is being charged.
• GREEN – Battery is fully charged.
BATTERY MAINTENANCE
Recharge the battery every three months when not in use.
If the scale is not used for an extended period of time,
remove the battery from the battery compartment to
avoid leakage. Store the battery in a sealed bag or box
in a dry, temperate environment.
PRIOR TO USE
• Fully charge battery before using scale for the first time.
• Let the scale warm up for 15 minutes before use.
WARNING! NEVER drop items onto the platform.
The scale is a sensitive precision instrument.
WARNING! Do not exceed the scale capacity
and overload the scale.
WARNING! Do not stack items on the scale
platform when the scale is not in use.
TURNING THE SCALE ON/OFF
Power switch is located on the right side of the scale.
CAUTION! Never power the scale on with weight
on the platform.
1. To turn the scale on, press the switch forward and
hold. The scale will show the version and capacity,
then enter a self-test mode. After completing the
self-test mode, the scale will enter the weighing
application mode.
2. To turn the scale off, press the switch backward and
hold.
CHANGE UNIT OF MEASURE
Press the
UNITS
key
Units
until the desired weighing unit
appears – lb, oz, kg or g.
NOTE: UNITS key operation is active in the weighing
and check weighing applications. The weighing
unit must be selected prior to the counting, check
counting and accumulation applications.
CALIBRATION
The scale is calibrated before leaving the manufacturer’s
facility. Vibrations in shipping may cause the scale to
fall out of calibration. Uline recommends re-verifying
calibration in the field before first use. You must use a
precision calibration weight to properly calibrate the scale.
The calibration weight must be at least 1/3 of the scale’s
capacity and cannot exceed the scale’s capacity.
NOTE: Calibration can be done in English
system weight (lb.) or Metric system weight (kg).
SINGLE POINT CALIBRATION (SPAN)
Single point calibration uses two calibration points: zero
and any weight between 1/3 of the scale’s capacity
and the full weight of the scale capacity. Refer to the
calibration weights charts below.
CALIBRATION WEIGHT (lb.)
Model #
1/3 WEIGHT (Minimum)
FULL CAPACITY (Maximum)
H-5819
4 lb.
11 lb.
H-5820
9 lb.
27 lb.
H-5821
19 lb.
55 lb.
H-5822
37 lb.
110 lb.
CALIBRATION WEIGHT (kg.)
Model #
1/3 WEIGHT (Minimum)
FULL CAPACITY (Maximum)
H-5819
2 kg.
5 kg.
H-5820
4 kg.
12 kg.
H-5821
9 kg.
25 kg.
H-5822
17 kg.
50 kg.
NOTE: Calibration weights are sold separately.
See uline.com for details.
1. Turn on the scale and press the
TARE
key
Tare
during
the self-test.
PO CAL
will be shown on the display.
2. Press the
ZERO
key
Zero
.
Lb
will be shown on the display.
a. If calibrating in lb., continue to step 3.
b. If calibrating in kg., press the
TARE
key
Tare
.
G
will be shown on the display.
3. Press the
ZERO
key
Zero
.
ULoAd
will be shown on the
display.
PO CAL
ULoAd