The TysonBio SUPERB results range between 20 to 600 mg/dL (1.1 to 33.3
mmol/L). If a “
HI
” message appears on your display, your TysonBio SUPERB
Meter has detected that your blood glucose level is higher than 600 mg/dL(33.3
mmol/L). It is suggested that you review your testing procedure and test again
with a new TysonBio SUPERB Test Strip to confirm the result.
If the same result
occurs, consult your healthcare professional immediately.
(Fig.7)
If a “
LO
” message appears on your display, your TysonBio SUPERB Meter has
detected that your blood glucose level is lower than 20 mg/dL (1.1 mmol/L). It is
suggested that you review your testing procedure and test again with a new
TysonBio SUPERB Test Strip to confirm the result.
If the same result occurs,
consult your healthcare professional immediately.
(Fig.8)
If no further tests are performed, the TysonBio SUPERB Meter will shut off
automatically after 4 minutes.
After finishing the test, we recommend using a tissue paper to remove TysonBio
SUPERB Test Strip from the TysonBio SUPERB Meter for proper disposal.
5.
6.
7.
M
S
The TysonBio SUPERB Meter automatically stores the 300 most recent blood
glucose results with date and time in its memory. It also provides you with a
consecutive 7/14/30-day average of your blood glucose test results.
Press the
M
button to turn on the TysonBio SUPERB
Meter. All segments of the LCD display will appear. The
date and time will appear first.
Press
S
button, the result of control solution will appear
on display.*
Press
S
button, the consecutive 7-day average will be
shown.*
Press
S
button, the consecutive 14-day average will be
shown.*
Press
S
button, the consecutive 30-day average will be
shown.*
1.
2.
3.
4.
5.
Note : * If there is no test results during this stage, it will
directly appear the most recent test result.