5.
6.
7.
After the “
“ symbol appears, press the
S
button for 3 sec to
switch control solution mode. The “ “ symbol indicates that
the Meter will mark your current test as a Control Solution
test. Now you are ready to apply the Control Solution.
Discard the first drop of Control Solution and squeeze a small
drop of Control Solution on a clean nonabsorbent surface
(such as a clean piece of wax paper).
Do not apply Control
Solution to the Test Strip directly from the bottle.
Bring the tip of the Test Strip to lightly touch the drop of
Control Solution. Control Solution is automatically pulled into
the strip through the tip. Hold until the meter beeps. The
meter will now being counting down from 8 to 1 and the
Control Solution test result will appear.
Caution: Please remember that in order to prevent
contamination, follow the above instructions for perform-
ing the Control Solution test.
mg/dL
AM
mg/dL
Comparing Control Solution Results
The TysonBio SUPERB is functioning properly if the result falls inside the speci-
fied range printed on the Test Strip vial. If the test result falls outside the specified
range, repeat the test.
Results falling outside the specified range may be caused by:
● Error in performing the test
● Expired or contaminated Control Solution
● Expired or contaminated TysonBio SUPERB Test Strips
● Improper coding of the TysonBio SUPERB Meter
● TysonBio SUPERB Meter malfunction
Note: The result will not be incorporated into the average in the Control Solution
test “ “ mode. DO NOT use the system if the problem persists.
Please call Tyson Bioresearch, Inc., Customer Service at +886-37-585998 or
website of www.tysonbio.com
M
S
27
28