KANGAROO JOEY

  

32

Enteral Feeding Pump

section X — specifications and symbols

specifications

Ul listing

Medical equipment

KANGAROO JOEY™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards 

in accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other 

specified hazards in accordance with CAN/CSA C22.2 No. 601.1

Type Infusion Device

Volumetric

Pumping Mechanism

Rotary Peristaltic

Pump sets

All sets operating on the KANGAROO ePump Enteral Feeding Pump will also operate on the KANGAROO JOEY Enteral Feeding Pump.

 

feeding formula Delivery Rate

1-400 mL/hr in 1 mL increments

 

feeding formula VTbD

1-3000 mL in 1 mL increments

bolus Volume

1-3000 mL in 1 mL increments

number of boluses

1-99

bolus Interval

1-24 hours in 1-hour increments

flushing solution Dose Range

10-500 mL in 1 mL increments

flushing solution Interval Range

1-24 hr in 1 hr increments

accuracy

7% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 15.24 cm (6”) ± 0.76 cm (0.3”) above the top of the closed 

pump door, at a room temperature of 22°C ± 2°C (72°F ± 3°F), using a new ePump feeding set for no longer than the recommended 

24 hours of maximum usage. Confidence intervals for accuracy are based upon those included in the ANSI/AAMI ID26-1992 American 

National Standard for Infusion Devices. Reported accuracy is based upon the time-based accuracy calculation in this standard, as 

applicable for enteral feeding. Since feeding accuracy of the KANGAROO JOEY Enteral Feeding Pump is based upon formula and 

not water, a 4% accuracy shift should be applied when testing the KANGAROO JOEY Enteral Feeding Pump using water. (Water will 

appear to overdeliver by 4%.)

Содержание Kendall Kangaroo Joey

Страница 1: ...Pumpa pro enter ln v ivu Provozn a servisn p ru ka Enter lis t pl l pumpa Kezel si s karbantart si k zik nyv Pompa ywienia dojelitowego Instrukcja obs ugi i serwisowania Enteral Besleme Pompas Kullan...

Страница 2: ...fication 4 Section IV Initial Setup Attaching the A C Adapter Power Cord 5 Battery Setup 6 Attaching Pole Clamp 6 SectionV Instructions for Use Quick Start 7 General Startup 8 Placement Mounting 8 A C...

Страница 3: ...Flushing 16 Run 17 KTO 18 Lock Screen Feature 18 Hold 19 Flush Now 20 More Options 20 Buzzer AlarmVolume 20 History 20 Language Selection 21 Continuous Intermittent Mode 21 BioTech Menu 21 Accessing...

Страница 4: ...s 32 Symbols on Pump and Accessories 35 Section XI Customer Service 36 Section XII Maintenance Top Door Enclosing Pump Set Loading Area 36 Battery Pack 37 Battery Door 37 Power Cord 37 Pole Clamp 37 S...

Страница 5: ...tTouch rubberized casing for easy grip and drop protection Quick Release pole clamp for easy detachment of pump Features MISTIC Magnetic Intelligent SetType Identification Connector Feed Only vs Feed...

Страница 6: ...wer Easily removable power cord Figure 1A KANGAROO JOEY Enteral Feeding Pump features front and rear views Menu Selections Menu Selection Buttons Graphic Display Front Door Back Battery Door Serial I...

Страница 7: ...see SectionVI Re certification of Performance For other integrity checks consult with a qualified Biomedical Technician or contact the manufacturer Section XI Customer Service 7 For service or for te...

Страница 8: ...nd the pump rotor 15 Do not attempt to rotate the valve stem inside the feeding set valve assembly Rotation of the valve stem can only be done by the pump while the valve assembly is properly loaded i...

Страница 9: ...of the pump See Figure 1A to locate port on pump NOTE It is important to make sure the power plug is pushed all the way into the pump for proper operation If the display screen shows the pump is runn...

Страница 10: ...18 hours of service before needing to be recharged Attaching Pole Clamp The KANGAROO JOEY Enteral Feeding Pump can be attached to a vertical pole using the pole clamp which is included with the pump a...

Страница 11: ...l 2 Open blue door enclosing pump set loading area 3 To load pump set refer to Figure 4 Grasp finger tab on valve and insert into front pocket 1 Grasp black ring retainer and wrap tubing around rotor...

Страница 12: ...MORE options menu 8 Press4RUN when ready to start The screen will displayRUNNING 9 To stop press4HOLD or press and hold the POWER button 182541 6 in 15 2 cm 6 in 15 2 cm Figure 4B Correct Fluid Level...

Страница 13: ...rough the MORE OPTIONS menu see Figure 10A Keep or Clear Prior Pump Settings Immediately after powering up the pump if the prior feed or flush settings were not cleared and the settings are not locked...

Страница 14: ...ot on the left Figure 6A 3 Grasp the MISTIC retainer end black ring retainer and wrap the tubing counterclockwise around the rotor Avoid overstretching the silicone tubing Figure 6B 4 Carefully pull t...

Страница 15: ...ethod using the hold to prime feature After a pump set has been loaded and the status line shows SETLOADED press4PRIME PUMP in the opening menu Figures 7A 7D to get one of the PRIMEPUMP menus shown in...

Страница 16: ...alve and then press4HOLDTO PRIME FEED until the feeding fluid has reached past the valve and down to the stepped end connector If the feeding line is primed first subsequent priming of the flushing li...

Страница 17: ...e may be selected through the Biotech Options button which may be accessed on the Biotech screen see page 22 The EZ Pump Mode setting will be stored in the machine pump memory and saved when the unit...

Страница 18: ...ogram the volume from 1 to 3000 mL in increments of 1 mL Press4ENTER top button to exit the menu If the FEEDVTBD is set to zero the pump will run until the supply is exhausted NOTE When a FEEDVTBD is...

Страница 19: ...rs from the ADJUSTBOLUS screen Figure 13A Press4BOLUS RATE in Figure 13A to define the rate of delivery Press the4buttons in Figure 13B to program the delivery rate from 1 to 400 in increments of 1 mL...

Страница 20: ...the time interval between starts of bolus Intermittent Mode Feeding Bolus Max Not Available in EZ Pump Mode The BOLUS MAX option can be used to deliver a bolus at a high rate similar to gravity feedi...

Страница 21: ...settings which exceed pump capabilities to deliver CAUTION The KANGAROO JOEY Enteral Feeding Pump flushes at a rate of 1960 mL hr 32 7 mL minute Use care when programming the flush volume so that it...

Страница 22: ...utton will increment the KTO time by 5 minute increments up to 30 minutes see Figure 15B After 30 is reached each additional button press will increment in larger time steps up to a maximum of240minut...

Страница 23: ...return to one of the opening menus shown in Figures 7A 7D thus allowing any setting to be modified If the pump set has been removed after going into HOLD the opening menu will show LOADASET as seen i...

Страница 24: ...USH NOW before the set volume has been delivered is desired press4FLUSH NOW and enter a volume of zero and press4ENTER The FLUSH NOW will stop immediately and the pump will return to the RUNNING mode...

Страница 25: ...nformation Press4DONE in the MORE OPTIONS menu to exit the menu BioTech Menu The BioTech screen will show technical information about the pump and has a lock feature to lock the Pump Settings against...

Страница 26: ...d again by pressing4LOCK SETTINGS to remove the check mark If the settings are locked and a change is attempted the screen will indicate PROGRAM LOCKED and will not allow access to the parameters Figu...

Страница 27: ...of the water column is 6 inches above the pump Figure 4B Make sure pump is plugged into A C power do not run test with battery power Program a Continuous Mode feed rate see below for suggested example...

Страница 28: ...power source before cleaning After cleaning do not connect to AC power source until pump and power cord are thoroughly dry A mild detergent should be used for general cleaning If necessary the pump ma...

Страница 29: ...Representative or call Customer Service for instruction Section VIII Battery Replacement See Figure 24 and do the following to replace the Battery Pack Power down the unit Loosen the two battery door...

Страница 30: ...AROO JOEY Enteral Feeding Pump display will show the error indicator icon with specific information regarding the error condition and will sound the alarm The following conditions will trigger error c...

Страница 31: ...otor current 5 Never got a ADC ISR for battery level 6 Never got a ADC ISR for battery temperature 7 Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag Empty checks 8 Never got a ADC ISR for...

Страница 32: ...ROTOR ERROR screen appears during RUNNING or PRIMING when the pump has detected an unusual operating condition for the rotor See Figure 27 The ROTOR ERROR generally results from a problem with the pu...

Страница 33: ...ient Check the line to find the occlusion causing the blockage If the error cannot be resolved check the valve pocket in the pump set loading area for dampness or dirt which may be a cause for a false...

Страница 34: ...n so it is flush with the back of the pump Figure 32 Battery Low error screen Indicates that the battery must be recharged immediately Feeding Complete Adjustable Audible Volume The FEEDING COMPLETE i...

Страница 35: ...status especially in darkened rooms A green light status indicates normal pump operation A yellow light status indicates an informational or warning situation for a low battery condition the pump in H...

Страница 36: ...f Boluses 1 99 Bolus Interval 1 24 hours in 1 hour increments Flushing Solution Dose Range 10 500 mL in 1 mL increments Flushing Solution Interval Range 1 24 hr in 1 hr increments Accuracy 7 or 0 5 mL...

Страница 37: ...er cord NominalVoltage 3 7V DC Max Current 1800 A Battery New fully charged Lithium Ion Battery Pack delivers 18 hrs at 125 mL hr feed rate Approximately 15 minutes prior to complete battery discharge...

Страница 38: ...75 R H non condensing StorageTemperature 0 50 C 32 122 F 95 R H non condensing Type of Protection Against Electrical Shock Class II Internally powered Equipment Degree of Protection Against Electrica...

Страница 39: ...luids CE Mark European Community Certification Power Button Buttons for Selecting Screen Prompts Power Source 5VDC 2 4 A Caution For indoor use only Functional Ground Do not use KANGAROO JOEY Enteral...

Страница 40: ...care Group LP Tyco Healthcare Canada 98 6 Faichney Drive 7300Trans Canadian Highway Watertown NY 13601 Pointe Claire QC H9R 1C7 1 800 448 0190 or 315 788 5246 1 877 664 8926 or 514 695 1220 U K Servic...

Страница 41: ...ed several times to restore optimal capacity Battery Door To replace Battery Door see Figure 36 and do the following 1 Place Battery Door over battery opening Make sure the rubber periphery seal on th...

Страница 42: ...ding Pump 383400 2 Main Door Blue transparent door Main Door Blue transparent door F32061 3 Battery Pack Battery Pack F010506 4 Battery Door Battery Door F31929 Power Cord with A C Adapter Power Cord...

Страница 43: ...he defective part or product with a comparable part or product repair the defective part or product or if neither replacement nor repair is reasonably available refund to Customer the purchase price f...

Страница 44: ...4 3 2002 Complies Conducted Disturbance Immunity EN60601 1 2 IEC 61000 4 6 2001 Complies Power Frequency Magnetic Field Immunity EN60601 1 2 IEC 61000 4 8 2001 Complies Voltage dips and sags Immunity...

Страница 45: ...manufacturing process The KANGAROO ePump AccuSystem feeding set manufacturing process parameters are tightly controlled atTyco Healthcare Group LP to minimize variation and produce consistent and rel...

Страница 46: ...sted per the pump s internal re certification protocol with the following results ManualTests Passed Failed BuzzerTest LEDTest BatteryTest MISTICTest AutoTest Inspector Name Signature Facility Address...

Отзывы: