Directions for Cleaning Rotor assembly
• Refer to General Cleaning Directions before starting.
• Open the blue door enclosing the pump set loading area.
• Loosen rotor set screw with 5/64” (2 mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid getting any
moisture in the rotor shaft opening.
• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
• Be sure all parts of rotor are completely dry before putting it back onto shaft.
• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
• Push rotor into place and tighten set screw (Do not over tighten).
Preventative Maintenance
This pump must be periodically re-certified to assure proper functioning and safety. The recommended service interval is every two
years. Re-certification may be done at the user’s Biomedical Engineering Department, an outside service, or by Tyco Healthcare
Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service in the US, call 1-800-448-0190. In Canada, call
1-877-664-TYCO(8926). Outside of the US and Canada, please contact your local Customer Service for more information.
If a pump malfunctions, please contact your Tyco Healthcare Group LP Representative or call Customer Service for instruction.
section VIII — battery Replacement
See Figure 24 and do the following to replace the Battery Pack:
• Power down the unit.
• Loosen the two battery door screws on the back of the pump.
• Remove battery door and disconnect wire harness making sure not to damage the watertight seal on the pump housing. A small
straight-bladed screw driver may be needed to disconnect the connector tab.
• Slide Battery Pack out of compartment.
• Slide a new Battery Pack into compartment with wire harness extending out.
• Push connector in until properly latched.
• Tuck wires into slot between connector and battery to insure wires are not pinched.
• Place battery door over battery opening. Make sure the rubber periphery seal on the pump housing fits into the groove in the
battery door before closing all the way. Avoid pinching the seal to retain a water-tight seal.
• Dispose of used battery cells promptly. Keep away from children.
• The battery is packed partially charged. Charge battery 6 hours prior to use off A/C power.
note:
Replacement Battery Packs are available from Tyco Healthcare Group LP (See Section XIII - Service Part Numbers).
To assure the safe and reliable operation of this pumping system, do not attempt to operate without an approved battery pack
installed in the pump.
note:
The KANGAROO JOEY Enteral Feeding Pump will not operate unless an approved KANGAROO JOEY Enteral Feeding Pump