![Tyco Kendall Kangaroo Joey Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-joey/kendall-kangaroo-joey_operating-and-service-manual_1186011012.webp)
KANGAROO JOEY
™
8
Enteral Feeding Pump
(
note:
display status line should read
SET LOADED
).
4.
suspend the pump set bag so that the top of the starting volume of formula is 6 inches above the pump (figure
4b).
5. To automatically prime the pump, press
4
PRIMe PUMP
and
4
aUTo PRIMe
. For Feed & Flush sets, auto priming will prime
both lines.
6. To directly control pump priming, press
4
PRIMe PUMP
and then press and hold the
HolD To PRIMe
button(s)
4
. If a Feed
& Flush Set is used, remember to first press
4
HolD To PRIMe flUsH
until the flushing fluid has reached the valve, and then
press
4
HolD To PRIMe feeD
until the feeding fluid has reached past the valve and down to the distal end connector.
7. Use buttons
4
on left side of LCD to set the feeding parameters. Set flushing parameters if a Feed & Flush Set is loaded. The
continuous or intermittent feed mode selection is made under the
MoRe
options menu.
8. Press
4
RUn
when ready to start. The screen will display
RUNNING
.
9. To stop, press
4
HolD
or press and hold the
PoWeR
button.
182541
6 in
(15.2 cm)
6 in
(15.2 cm)
figure 4b. Correct fluid level.
figure 4a. Installation of Pump set. (see figure 6 for detailed view.)
WaRnInG: Do noT oVeRsTReTCH TUbe.
1
3
2
black Ring
Retainer
(MIsTIC)
finger
Tab
Valve
Rotor
General startup
Placement/Mounting
There are two recommended placement methods for the KANGAROO JOEY Enteral Feeding Pump.
• Attached to a vertical IV pole via the pole clamp peripheral device, included with the pump.
• Placed on any stable surface.
a/C Power operation
Plug the KANGAROO JOEY Enteral Feeding Pump into an A/C power outlet for normal operation. When the pump is not plugged in,
or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
battery Power operation
If A/C power is not available to the pump, or if A/C power is interrupted, the pump will automatically operate on backup battery
power. The built-in battery recharges any time the pump is plugged into an outlet.