![Tyco Kendall Kangaroo Joey Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-joey/kendall-kangaroo-joey_operating-and-service-manual_1186011017.webp)
KANGAROO JOEY
™
13
Enteral Feeding Pump
Turning on
eZ PUMP MoDe
will automatically force the unit out of the
InTeRMITTenT (bolUs feeD)
mode if set. Although the
Continuous Mode feed VTBD will be set to 0 while
eZ PUMP MoDe
is enabled, the previous value of the
feed VTbD
will be stored
in the pump memory and recalled when “
eZ PUMP MoDe
” is deselected. The user can change the Feed Rate by normal means. All
other user settings, including those not compatible with the
eZ PUMP MoDe
(
bolUs VTbD
,
nUMbeR of bolUses
,
sUPeR
bolUs
, etc), that were set before the pump was put into the
eZ PUMP MoDe
will be ignored and invisible while in
eZ PUMP
MoDe
but remain stored in the pump memory for when
eZ PUMP MoDe
is deselected.
eZ PUMP MoDe
essentially has no effect
on
CleaR seTTInGs
; pressing
CleaR seTTInGs
while
eZ PUMP MoDe
is enabled will clear all Feed and Flush settings and
totalizers as normal, not just those that apply to
eZ PUMP MoDe
. Selecting the
CleaR seTTInGs
option on power up will not
clear
biotech options
such as
eZ PUMP MoDe
or
loCK seTTInGs
.
If the user selects the
aDJUsT feeD
option from either the
loaD a seT
or
seT loaDeD
screens, control will pass directly to and
from the Enter Feed Rate screen and bypass the
aDJUsT feeD
screen since without the ability to adjust VTBD in
eZ PUMP MoDe
,
the Adjust Feed screen is redundant.
selecting eZ Pump Mode
eZ PUMP MoDe
may be selected through the
biotech options
button which may be accessed on the Biotech screen (see page
22). The
eZ PUMP MoDe
setting will be stored in the machine pump memory and saved when the unit is shut off.
noTe:
If the user inserts a Feed/Flush set while the
eZ PUMP MoDe
is enabled,
eZ PUMP MoDe
will be cancelled and the unit
will immediately revert back to the normal mode. If
eZ PUMP MoDe
is desired, the user must then re-enable
eZ PUMP MoDe
through the
biotech options
screen.
When enabled and not in
CeRTIfICaTIon MoDe
, a small
eZ
icon will be visible on the bottom of the screen near the battery icon.
However, it will not appear in the
PoWeR DoWn
,
sYsTeM eRRoR
, or
loW baTTeRY sCReens
.
figure 9.
selecting Continuous or Intermittent feed Mode
The KANGAROO JOEY Enteral Feeding Pump can be programmed to feed continuously or intermittently.
The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has
been delivered or until the supply has been exhausted. The Intermittent feeding mode delivers boluses of enteral nutrition at
programmed time intervals. The bolus volume and feed rate are also programmed.
To select continuous feed mode or intermittent feed mode, press
4
MoRe
on the opening menu, shown in Figures 5A-5B or Figures
7A-7D, to access the
MORE OPTIONS
menu, Figure 10A.
The
MoRe
option is available either before a pump set has been loaded (Figures 5A-5B) or after the pump set is loaded (Figures 7A-
7D). See subsection “More Options” for information on the other options in the
MORE OPTIONS
menu.