![Tyco Kendall Kangaroo Joey Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-joey/kendall-kangaroo-joey_operating-and-service-manual_1186011024.webp)
KANGAROO JOEY
™
20
Enteral Feeding Pump
flush now
(Not Available in EZ Pump Mode)
During the
RUNNING
mode, the pump can be diverted to immediately perform a flush (flush on demand) when a feed/flush set
is loaded by pressing
4
FLUSH NOW
in the
RUNNING
menu, input a desired flush volume, then press
4
enTeR
. See Figure 15A.
When
4
flUsH noW
is pressed, the screen will prompt for the volume to be immediately flushed, and will show the previously
programmed flush volume as default but may be adjusted to the users desired volume. The screen will appear as the standard
Flush input screen as shown in Figure 12B (or 14B). Setting the
flUsH noW
flushing volume is temporary and will not change the
main/periodic flushing volume previously programmed. If
4
flUsH noW
is pressed by accident; the user may either press the cancel
button or allow the screen to return to the
RUNNING
screen after 10 seconds, if there is no input for the flush volume.
If stopping a
flUsH noW
before the set volume has been delivered is desired, press
4
flUsH noW
and enter a volume of zero and
press
4
enTeR
. The
flUsH noW
will stop immediately and the pump will return to the
RUNNING
mode.
More options
Press
4
MoRe
in one of the opening menus, Figures 5A-5B or Figures 7A-7D, to access the
MORE OPTIONS
screen, Figure 19.
figure 19.
MORE
OPTIONS
menu.
buzzer
Press
4
bUZZeR
in the
MORE OPTIONS
menu to access the
BUZZER
screen and increase or decrease the alarm buzzer volume. The
alarm will be audible as the volume setting is changed. This sets the volume level for any warning or error that may occur with the
pump. Critical errors of System Error and Pump Set Dislodged will always use the maximum volume setting.
History
(Not Available in EZ Pump Mode)
Press
4
HIsToRY
in the
MORE OPTIONS
menu to access the
HISTORY
screen. Press the
4
buttons in Figure 20 to set the time history
of interest. The totals of
feeD ml
and
flUsH ml
will be shown for the requested history time.
WaRnInG:
A maximum of 72 hours of past history can be retrieved, excluding data from the current hour. NOTE: The 72-hour
history includes time when the pump is powered OFF.