background image

 

 

 

 

Wire and Harnessing Products

 

Engineering Specification 
 

 

 No: 

ES 61402

  

 Rev: 

 

Date:  July 9, 1991 

 

Page:  12 of 16 

 

If this document is printed it becomes uncontrolled.  Check for the latest revision.

 

4.0 

Calibration Verification Test 

4.1 Objective 

This procedure is performed to verify whether the Waffle Iron II control system is operating 
correctly and the heating temperatures are within the specified ranges. Results of this test will 
determine if the heating tool can be used, or if it must be recalibrated by an authorized 
Raychem representative. No adjustments are to be performed. 

4.2 Tools 

Required 

o

Test platen: 
a.  For early model Waffle Iron II's without thermocouple connection on back, the 

CE-1406800 calibration platen is required. 

 

b.  For late model Waffle Iron II's with thermocouple connection on back of unit, any 

platen set in good condition or a CE-1406800 calibration platen can be used. 

 

o

Digital thermometer with peak hold feature, or strip chart recorder; must be 
compatible with type K thermocouple. 

 

o

Stop watch 
 

4.4 Test 

Procedure 

1. Install test platen: 

o

For early model Waffle Iron II's without thermocouple connection  on  back  of  unit, 
remove the production platen and install the CE-1406800 calibration platen. The 
calibration platen locks into the lower heat sink the same as a production platen. 

 

o

For late models with thermocouple connection, install the normal production platen 
set in the Waffle Iron II. (The CE-1406800 calibration platen may also be used.) 

 

2.  Close and latch the upper heat sink. 
3.  Connect the temperature recording device to the thermocouple. 

 

o

If the Waffle Iron II is an early model without a thermocouple connection, connect 
the thermocouple lead of the CE-1406800   calibration platen into the temperature 
recording device. 

o

If the Waffle Iron II is a late model and has a thermocouple connection, connect the 
thermocouple lead (supplied) between the temperature recording device and the 
thermocouple connection on the back of the unit. 

 

4.  Turn the Waffle Iron II on. 

Содержание CE-1404200

Страница 1: ...ES 61402 Rev D Date July 9 1991 Page 1 of 16 If this document is printed it becomes uncontrolled Check for the latest revision RAYCHEM INTERCONNECTION SYSTEMS DIVISION ENGINEERING STANDARD ES 61402 OP...

Страница 2: ...ducts Engineering Specification No ES 61402 Rev D Date July 9 1991 Page 2 of 16 If this document is printed it becomes uncontrolled Check for the latest revision OPERATING AND MAINTENANCE MANUAL WAFFL...

Страница 3: ...WARNING Do not touch infrared lamps They may be hot and can cause burns WARNING In keeping with good industrial hygienic practice adequate ventilation must be maintained whenever plastic materials ar...

Страница 4: ...cable CTM transitions for splicing flat conductor cable to round wire MTC50 and MTC100 connector wafers for flat conductor cable and SEM connector wafers for flat conductor cable The part to be termin...

Страница 5: ...ime period selected After the timed heating period the heat lamps and HEATING indicator light go out the COOLING indicator light illuminates and the blower comes on to cool the platens The cooling cyc...

Страница 6: ...0 120 V 100 120 VAC 50 60 Hz 20 A Model CE 1404300 240 V 200 240 VAC 50 60 Hz 10 A Overall Cycle Time 3 4 Minutes Temperature Per TEMP control dial Setting 0 through 9 Set Point Range Time at Temperat...

Страница 7: ...installation instructions Step 2 Before installation or after cleaning spray or wipe platens with light coat of release agent such as Miller Stephenson MS 122 or Axel MoldWiz o Coat only the side of t...

Страница 8: ...r the latest revision Step 5 Remove install platens as required o Platens are locked into heat sinks by a spring loaded retainer o Identification markings face the operator o Top platen has one keying...

Страница 9: ...tive Step 3 Press the START button to initiate a heating cycle to warm the tool and verify its operation o The TEMP and TIME controls can be at any setting for this cycle o The green READY ERROR light...

Страница 10: ...3 Load the assembly to be terminated onto the lower platen of the heating tool o Position CTMs so that the top surface faces upward and the CTM is centered in the platen o Position MTC and SEM connect...

Страница 11: ...flash until the heat sink is closed and the START button is pressed Step 8 Allow the termination cycle to continue until the green READY ERROR light comes on o After the heating portion of the cycle i...

Страница 12: ...gital thermometer with peak hold feature or strip chart recorder must be compatible with type K thermocouple o Stop watch 4 4 Test Procedure 1 Install test platen o For early model Waffle Iron II s wi...

Страница 13: ...uirements in the temperature table below The maximum temperature reached during the 10 second window must fall within the applicable range from the temperature table IMPORTANT A numerical correction f...

Страница 14: ...ty using a soft cloth or tissue moistened with trichloroethane or isopropyl alcohol Do not touch lamps directly with fingers 5 2 Preventing Parts From Sticking Before installation or after cleaning sp...

Страница 15: ...ller fault Contact Raychem field service Flashing green Upper heat sink not Close and latch upper READY ERROR closed and latched heat sink light Operator error Previous termination Remove previous ter...

Страница 16: ...t becomes uncontrolled Check for the latest revision 5 5 Replacement Parts List DESCRIPTION PART NUMBER Infrared lamp top or bottom 966723 Reflector top or bottom 977842 Latch stud for heat sink 96624...

Отзывы: